Varian Medical Systems, Inc. Oncology Systems: GammaMed Plastic Needle with Mandrin; 2.0mm diameter, length 200 mm. Remote controlled radionuclide applicator system. Varian Medical Systems, Inc. Mfg by Varian Medical Systems Haan GmbH, Haan, Germany Product Usage: The Interstitial Plastic Needles with 2 mm diameter are designed for interstitial treatments in areas such as head and neck, gynecology, breast or prostate. In order to avoid bending and to facilitate the introduction of the needles each needle is supplied with a Mandrin
- Recall number
- Z-2992-2011
- Recalling firm
- Varian Medical Systems, Inc. Oncology Systems
- Product
- GammaMed Plastic Needle with Mandrin; 2.0mm diameter, length 200 mm. Remote controlled radionuclide applicator system. Varian Medical Systems, Inc. Mfg by Varian Medical Systems Haan GmbH, Haan, Germany Product Usage: The Interstitial Plastic Needles with 2 mm diameter are designed for interstitial treatments in areas such as head and neck, gynecology, breast or prostate. In order to avoid bending and to facilitate the introduction of the needles each needle is supplied with a Mandrin
- Status
- Terminated
- Initiated
- 2011-07-13
- Posted
- 2011-08-10
- Terminated
- 2012-06-20
- Reason
- Needle tip may separate from the needle shaft such that the tip may detach in the patient upon withdrawal.
- Root cause
- Device Design
- Action
- Varian sent an "Urgent Medical Device Correction/Urgent Field Safety Notice" letter dated July 13, 2011 to all possibly affected customers. This notification supplements and modifies the previous "Customer Technical Bulletins" originally sent in September 2008. The initial Customer Technical Bulletins informed customers of the possibility that needle tips cans crack and separate. It also reminded users that the product has a limited life and to check needles prior to use as per instructions. Current correction notice letter instruct customers to stop using any plastic needles in inventory with lot numbers C01 to D01 and those with lot numbers starting with H or I. The letter also request that the recalled products be returned to Varian BrachyTherapy for replacement. The letters include information as to where the lot number is located on the device. All appropriate personnel should be notified of the correction notice. Contact information is provided for all geographical areas. USA contact for questions or concerns is 1-800-360-7909.
- Quantity
- 601 - all sizes.
- Distribution
- Worldwide Distribution - USA (nationwide) and the countries of: Argentina, Austria, Bangladesh, Belgium, Brazil, Canada, Columbia, Cyprus, Czech Republic Denmark, Dominican Republic, Estonia, France, Germany, Greece, Hungary, India, Japan, Kazahastan, Kuwait, Lebanon, Macedonia, Mexico, Moldova, Nepal New Zealand, Norway, Pakistan, Philippines, Poland, Russia, Serbia, Singapore, Slovenia, South Africa, South Korea, Spain Sweden, Switzerland, Taiwan, Thailand, Trinidad, Tobago, Turkey, Ukraine, United Arab Emirates Uruguay, United Kingdom, and Venezuela.
- Lot, serial or model codes
- Part numbers GM11007570; All lots C01 to D01 and lots H and I
- 510(k) numbers
- ["K983436"]
- PMA numbers
- not on file
- Product code
- JAQ
- Firm FEI
- 2916710
- Firm city
- Palo Alto
- Firm state
- CA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28