FDA Medical Device Recalls (openFDA)
American Contract Systems Inc: American Contract Systems medical procedure kits and trays labeled as: 1) ANTERIOR CERVICAL PACK - 205988, Model Item Number MHAC65AH; 2) SPINE PACK - 206013, Model Item Number MHSP99AF.
- Recall number
- Z-2990-2026
- Recalling firm
- American Contract Systems Inc
- Product
- American Contract Systems medical procedure kits and trays labeled as: 1) ANTERIOR CERVICAL PACK - 205988, Model Item Number MHAC65AH; 2) SPINE PACK - 206013, Model Item Number MHSP99AF.
- Status
- Open, Classified
- Initiated
- 2026-05-19
- Posted
- 2026-08-20
- Terminated
- not on file
- Reason
- The recall notice was initiated due to a Medicom Urgent Medical Device Recall. Medicom is recalling specific lots of Ritmed Neurological Sponges Non-Sterile Models 55992C to 55998C due to potential for unexpected variability of endotoxin levels on the products.
- Root cause
- Nonconforming Material/Component
- Action
- American Contract Systems issued an "URGENT: MEDICAL DEVICE FIELD CORRECTION" on 5/19/2026 with a revision issued on 6/4/2026 via email. The notice explained the issue, risks to health, and requested the consignee affix a recall label from the template provided to the exterior of the kit. The label states: "URGENT: RECALL NOTIFICATION, Prior to use, remove & discard the Medicos Ritmed Neurological Sponge per ACS Recall FA-2026-034." Customers/end-users receiving the product through a distributor were directed to contact the distributor or sales representative with any questions. Complete the Recall Response Form. Distributors were directed to directed to ensure all end-users are appropriately notified and maintain records of effectiveness. If you have any questions, contact the Regulatory Affairs Manager, Global Products by email: GM-OMRA_RECALLS@owens-minor.com or phone at: 470.280.4277 .
- Quantity
- 160 kits
- Distribution
- US distribution to states of: MN, MO, IA
- Lot, serial or model codes
- Model/Item Number MHAC65AH, UDI/DI 191072211910, Lot Number 42-8606211; Model/Item Number MHAC65AH, UDI/DI 191072211910, Lot Number 42-8687611; Model/Item Number MHAC65AH, UDI/DI 191072211910, Lot Number 42-8847511; Model/Item Number MHSP99AF, UDI/DI 191072214782, Lot Number 42-8423411; Model/Item Number MHSP99AF, UDI/DI 191072214782, Lot Number 42-8470411; Model/Item Number MHSP99AF, UDI/DI 191072214782, Lot Number 42-8527611; Model/Item Number MHSP99AF, UDI/DI 191072214782, Lot Number 42-8601311; Model/Item Number MHSP99AF, UDI/DI 191072214782, Lot Number 42-8731511; Model/Item Number MHSP99AF, UDI/DI 191072214782, Lot Number 42-8795311.
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- OJH
- Firm FEI
- 3008492337
- Firm city
- Tiffin
- Firm state
- OH
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28