FDA Medical Device Recalls (openFDA)
MEDLINE INDUSTRIES, LP - Northfield: Medline Convenience kits, labeled as: SPINAL BLOCK 22G QUINCKE 5S BU, Pack Number DYNJRA9026
- Recall number
- Z-2989-2024
- Recalling firm
- MEDLINE INDUSTRIES, LP - Northfield
- Product
- Medline Convenience kits, labeled as: SPINAL BLOCK 22G QUINCKE 5S BU, Pack Number DYNJRA9026
- Status
- Open, Classified
- Initiated
- 2024-04-08
- Posted
- 2024-09-06
- Terminated
- not on file
- Reason
- Medical convenience kits contain plastic syringes affected by the FDA Safety Alert issued on March 19, 2024. Leaks, breakage, and/ other quality issues have been identified that may pose a risk to patient health.
- Root cause
- Under Investigation by firm
- Action
- Medline Industries LP issued a recall notice to its consignees on 04/08/2024 via letter delivered via USPS and email. The notice explained the issue, provided a link to for information pertaining to the risk involved https://www.fda.gov/medical-devices/safety-communications/update-evaluating-plastic-syringes-made-china-potential-device-failures-fda-safety-communication) and request these syringes within the convenience be discarded, except for circumstances where no alternative syringes are readily available. If these syringes must be used, closely monitor for leaks, breakage, and any other problems, and report any quality issues to Medline and the FDA. Distributors were directed to notify their customers. The firm issued a second recall notice (an expansion) on 05/02/2024 using the same methods.
- Quantity
- not on file
- Distribution
- Worldwide distribution.
- Lot, serial or model codes
- DYNJRA9026 , Lot Number 22EBN191 ; DYNJRA9026 , Lot Number 22LBC356 ; DYNJRA9026 , Lot Number 23ABN403 ; DYNJRA9026 , Lot Number 23CBP475 ; DYNJRA9026 , Lot Number 23DBG934 ; DYNJRA9026 , Lot Number 23EBH050 ; DYNJRA9026 , Lot Number 23JBS641 ; DYNJRA9026 , Lot Number 23KBJ179 ; DYNJRA9026 , Lot Number 23LBU686 ; DYNJRA9026 , Lot Number 19ADC324 ; DYNJRA9026 , Lot Number 19ADD208 ; DYNJRA9026 , Lot Number 19LDA087 ; DYNJRA9026 , Lot Number 19LDB132 ; DYNJRA9026 , Lot Number 20ADC808 ; DYNJRA9026 , Lot Number 20BDA346 ; DYNJRA9026 , Lot Number 20BDB555 ; DYNJRA9026 , Lot Number 20CDB731 ; DYNJRA9026 , Lot Number 20EDC526 ; DYNJRA9026 , Lot Number 20GDA908 ; DYNJRA9026 , Lot Number 20IDA046 ; DYNJRA9026 , Lot Number 20IDA621 ; DYNJRA9026 , Lot Number 20LDB113 ; DYNJRA9026 , Lot Number 20LDB323 ; DYNJRA9026 , Lot Number 21BDA498 ; DYNJRA9026 , Lot Number 21DDA458 ; DYNJRA9026 , Lot Number 21DDA663 ; DYNJRA9026 , Lot Number 21HDC456 ; DYNJRA9026 , Lot Number 21JDB091 ; DYNJRA9026 , Lot Number 21KDC632 ; DYNJRA9026 , Lot Number 21LDA581 ; DYNJRA9026 , Lot Number 21LDB014 ; DYNJRA9026 , Lot Number 22BDA209 ; DYNJRA9026 , Lot Number 22BDC191 ; DYNJRA9026 , Lot Number 22CDA655 ; DYNJRA9026 , Lot Number 22CDC010 ; DYNJRA9026 , Lot Number 22ILB005 ; DYNJRA9026 , Lot Number 22KLA115
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- OJH
- Firm FEI
- 1417592
- Firm city
- Northfield
- Firm state
- IL
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28