Skip to the content

FDA Medical Device Recalls (openFDA)

ConvaTec, Inc: Esteem Body Soft Convex Urostomy Pouch Model/Catalog Numbers: 423703, 423704, 423705, 423707, 423708, 423709, 423710, 423712, 423713, 423714, 423715, 423716, 423717, 423718, 423721, 423722, 423723, 423725, 423726, 423727.

Recall number
Z-2986-2026
Recalling firm
ConvaTec, Inc
Product
Esteem Body Soft Convex Urostomy Pouch Model/Catalog Numbers: 423703, 423704, 423705, 423707, 423708, 423709, 423710, 423712, 423713, 423714, 423715, 423716, 423717, 423718, 423721, 423722, 423723, 423725, 423726, 423727.
Status
Open, Classified
Initiated
2026-07-06
Posted
2026-08-19
Terminated
not on file
Reason
Urostomy Pouches may leak from cap during use.
Root cause
Material/Component Contamination
Action
On July 20, 2026 Convatec issued a Urgent Medical Device Recall notification to affected consignees via Email. Convatec asked consignees to take the following actions: 1. Examine your inventory and identify all affected product lots using the lot numbers provided in the communication. 2a. (end-user) If you identify affected product, discontinue use of the affected device and contact your supplier, healthcare provider, or Convatec representative to obtain replacement product as soon as possible. If replacement product is not immediately available and continued use is necessary, follow the interim precautionary instructions. 2b. (distributor) Immediately quarantine and discontinue distribution, sale, transfer, and use of all affected product to prevent further distribution or use. 3. Inform all appropriate personnel within your organization. 4.If you have further distributed, transferred, or supplied any affected product, you must immediately notify all consignees and customers who may have received the affected product. Provide them with a copy of this recall notification and instruct them to take the actions described in this notice. 5. Please confirm receipt of this notice and your compliance with the required actions by completing and returning the enclosed acknowledgement form to us.customerservice@CONVATEC.COM no later than by August 07, 2026.
Quantity
617850 units
Distribution
US Nationwide distribution including in the states of AR, AZ, CA, CO, DC, DE, FL, GA, IA, IL, IN, LA, MA, MD, MI, MN, MO, MS, NC, NJ, NY, OH, OR, PA, RI, SC, TN, TX, UT, VA, WA.
Lot, serial or model codes
Lot Code: Product Name : Esteem Body" Urostomy Pouches UDI-DI: 00768455204526, 00768455204533, 00768455204540, 00768455204564, 00768455204571, 00768455204588, 00768455204595, 00768455204618, 00768455204625, 00768455204632, 00768455204649, 00768455204656, 00768455204663, 00768455204670, 00768455204700, 00768455204717, 00768455204724, 00768455204748, 00768455204755, 00768455204762 Lot / Serial Numbers for US market : 5H0475, 5J00257, 5J00475, 5J02234, 5J02340, 5J02361, 5J02732, 5K00013, 5K00338, 5K00713, 5K00338, 5K00713, 5K01320, 5K01500, 5K01541, 5K02079, 5K02104, 5K03003, 5K03008, 5L00231, 5L00444, 5L01041, 5L01042, 5L01162, 5L01935, 5L04677, 5L04679, 6A00067, 6A00218, 6A00219, 6¿01521, 6A01717 and 6A01964 Lot / Serial Numbers for global : 5J00020 , 5H04775 , 5J00257 , 5J00475 , 5J00476 , 5J00904 , 5J01425 , 5J01426 , 5J01512 , 5J01903 , 5J02218 , 5J02235 , 5J02340 , 5J02234 , 5J00903 , 5J02361 , 5J02362 , 5J01970 , 5J02732 , 5J04994 , 5J04995 , 5J05281 , 5K00712 , 5K00013 , 5K01500 , 5K01541 , 5K02080 , 5J05497 , 5K01320 , 5K02079 , 5K03003 , 5K00012 , 5K03008 , 5K01321 , 5K00338 , 5K03004 , 5K00713 , 5K02104 , 5K02095 , 5K06011 , 5K06313 , 5L00050 , 5L00231 , 5L00444 , 5K06686 , 5L00445 , 5L00691 , 5L01041 , 5L01042 , 5L01162 , 5L01508 , 5L01935 , 5L02306 , 5L02524 , 5L02796 , 5L03300 , 5L00948 , 5L04679 , 5L04677 , 5L05414 , 5L05415 , 6A00216 , 6A00217 , 6A00219 , 6A01188 , 6A00218 , 6A01717 , 6A01189 , 6A01521 , 6A00067 , 6A01522 , 6A01962 , 6A01964 , 6A01967 , 6A01965 , 6A01966 , 6A01963 , 6A02696
510(k) numbers
not on file
PMA numbers
not on file
Product code
EXG
Firm FEI
3011987967
Firm city
Greensboro
Firm state
NC
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register