FDA Medical Device Recalls (openFDA)
MEDLINE INDUSTRIES, LP - Northfield: Medline Convenience kits, labeled as: 1) DYNJ39451D , Pack Number HOOD PACK ; 2) DYNJ59573A , Pack Number ICU RESPIRATORY BRONCH PACK ; 3) DYNJ80281A , Pack Number OSC BRONCH PACK
- Recall number
- Z-2985-2024
- Recalling firm
- MEDLINE INDUSTRIES, LP - Northfield
- Product
- Medline Convenience kits, labeled as: 1) DYNJ39451D , Pack Number HOOD PACK ; 2) DYNJ59573A , Pack Number ICU RESPIRATORY BRONCH PACK ; 3) DYNJ80281A , Pack Number OSC BRONCH PACK
- Status
- Open, Classified
- Initiated
- 2024-04-08
- Posted
- 2024-09-06
- Terminated
- not on file
- Reason
- Medical convenience kits contain plastic syringes affected by the FDA Safety Alert issued on March 19, 2024. Leaks, breakage, and/ other quality issues have been identified that may pose a risk to patient health.
- Root cause
- Under Investigation by firm
- Action
- Medline Industries LP issued a recall notice to its consignees on 04/08/2024 via letter delivered via USPS and email. The notice explained the issue, provided a link to for information pertaining to the risk involved https://www.fda.gov/medical-devices/safety-communications/update-evaluating-plastic-syringes-made-china-potential-device-failures-fda-safety-communication) and request these syringes within the convenience be discarded, except for circumstances where no alternative syringes are readily available. If these syringes must be used, closely monitor for leaks, breakage, and any other problems, and report any quality issues to Medline and the FDA. Distributors were directed to notify their customers. The firm issued a second recall notice (an expansion) on 05/02/2024 using the same methods.
- Quantity
- not on file
- Distribution
- Worldwide distribution.
- Lot, serial or model codes
- DYNJ59573A , Lot Number 23IBP764 ; DYNJ59573A , Lot Number 23JBM050 ; DYNJ59573A , Lot Number 23KBM610 ; DYNJ59573A , Lot Number 24ABC779 ; DYNJ39451D , Lot Number 22BDB647 ; DYNJ39451D , Lot Number 22CDA099 ; DYNJ39451D , Lot Number 22LDA195 ; DYNJ39451D , Lot Number 23DDB501 ; DYNJ39451D , Lot Number 23DDB666 ; DYNJ39451D , Lot Number 23EDA821 ; DYNJ80281A , Lot Number 23ADA194 ; DYNJ80281A , Lot Number 23DDA881 ; DYNJ80281A , Lot Number 23HDA513 ; DYNJ80281A , Lot Number 23JDC172 ; DYNJ80281A , Lot Number 23LDB860 ; DYNJ80281A , Lot Number 24ADC258 ; DYNJ80281A , Lot Number 24CDA973
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- OFR
- Firm FEI
- 1417592
- Firm city
- Northfield
- Firm state
- IL
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28