Skip to the content

FDA Medical Device Recalls (openFDA)

Medtronic Sofamor Danek USA Inc: ATLANTIS(R) Anterior Cervical Plate System Screw Caddy; Self Tapping Variable Screw Caddy (176-513). Medtronic Sofamor Danek USA, Inc. This instrument is a precision device which incorporates a measuring function.

Recall number
Z-2984-2011
Recalling firm
Medtronic Sofamor Danek USA Inc
Product
ATLANTIS(R) Anterior Cervical Plate System Screw Caddy; Self Tapping Variable Screw Caddy (176-513). Medtronic Sofamor Danek USA, Inc. This instrument is a precision device which incorporates a measuring function.
Status
Terminated
Initiated
2011-07-06
Posted
2011-08-10
Terminated
2013-04-26
Reason
Lines on the screw gauge may not placed on the caddy properly which could result in improper measurement of screws.
Root cause
Process control
Action
Medtronic Sofamor Danek USA, Inc. sent "Voluntary Recall - Medtronic" letters on July, 2011 to all affected customers. The letter included; description of product and problem, and asked customers to stop using the product. Customers are asked to contact their Sales Representatives for replacement. For question on this recall call Medtronics Global Quality Department at (800) 876-3133.
Quantity
873 units
Distribution
Worldwide Distribution: Nationwide distribution including the states of MN, OK, IL, AK, CA, PA, UT, LA, SC, NY, MI, GA, WI, KY, AL, MO, AZ, TN, TX, MA, NE, OH, FL, CO, MT, CT, VA, NC, NH, WA, WV, WY, IN, IA, KS, MT, ME, DE, DC, NJ, MD, MS, NV, HI, ID, ND, and AR; and the country of Uruguay.
Lot, serial or model codes
**All lots.
510(k) numbers
not on file
PMA numbers
not on file
Product code
XLH
Firm FEI
1000477302
Firm city
Memphis
Firm state
TN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register