Skip to the content

FDA Medical Device Recalls (openFDA)

Roche Diagnostics Operations, Inc.: Elecsys Anti-TPO Assay on cobas e analyzers 411, 601, 602, Catalog Number 06368590190

Recall number
Z-2981-2020
Recalling firm
Roche Diagnostics Operations, Inc.
Product
Elecsys Anti-TPO Assay on cobas e analyzers 411, 601, 602, Catalog Number 06368590190
Status
Terminated
Initiated
2020-08-20
Posted
not on file
Terminated
2021-10-22
Reason
Recent internal verification studies found the current RF interference claim to no longer be met for the assay on the cobas e 411, 601, and 602 analyzers. Therefore, the claim has been updated.
Root cause
Nonconforming Material/Component
Action
Urgent Medical Device Correction notification letters dated 8/20/20 were distributed to customers. Actions Required " Be aware of the new RF interference claims listed in this Urgent Medical Device Correction (UMDC). This UMDC serves as labeling until the updated method sheet (V3), anticipated in September 2020, is available. " Please consult with the physician or pathologist at your facility to determine specific clinical implications for your patients. Customers do not need to perform an additional RF test unless otherwise determined in consultation with the physician or pathologist at your facility. " Complete the attached fax back form (TP-01048) and fax or email it according to the instructions on the form. " File this UMDC for future reference. Please consult with the physician or pathologist at your facility to determine specific clinical implications for your patients. The Elecsys Anti-TPO method sheet is being updated with the new claim for Rheumatoid Factor (RF) interference. Until the updated method sheet (V3) is available, this Urgent Medical Device Correction serves as your labeling. When these method sheets are updated, they will be available on the diagnostics.roche.com website and on the cobas e-library. Please contact the Roche Support Network Customer Support Center, 24 hours a day, seven days a week at 1-800-428-2336 if you have questions about the information contained in this UMDC.
Quantity
11,870 reagent kits
Distribution
US Nationwide
Lot, serial or model codes
All current and upcoming lots
510(k) numbers
K051890
PMA numbers
not on file
Product code
JZO
Firm FEI
1823260
Firm city
Indianapolis
Firm state
IN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this source