FDA Medical Device Recalls (openFDA)
Toshiba American Medical Systems Inc: Aplio XG model SSA-790A Aplio XG (SSA-790A) is a premium-class diagnostic ultrasound system incorporating state-of-the-art technologies. It is provided with the highest level of image quality and operability (reflecting the XG (next Generation) in its name), and can be used with ease for research as well as for routine clinical studies.
- Recall number
- Z-2981-2011
- Recalling firm
- Toshiba American Medical Systems Inc
- Product
- Aplio XG model SSA-790A Aplio XG (SSA-790A) is a premium-class diagnostic ultrasound system incorporating state-of-the-art technologies. It is provided with the highest level of image quality and operability (reflecting the XG (next Generation) in its name), and can be used with ease for research as well as for routine clinical studies.
- Status
- Terminated
- Initiated
- 2010-06-02
- Posted
- 2011-08-10
- Terminated
- 2012-04-20
- Reason
- Toshiba America Medical Systems Inc (TAMS) is initiating a field correction on the the Aplio XG model SSA-790A because it did not meet the requirement of clause 15b, IEC60601-1 which requires that the residual voltage shall not exceed 60V after 1 second from taking off the AC plug.
- Root cause
- Device Design
- Action
- Toshiba America Medical Systems (TAMS) sent an "URGENT:MEDICAL DEVICE CORRECTION' letter dated June 2, 2010 to all affected customers. The letter describes the product, problem, and actions to be taken by the customers. Customers were instructed to be careful not to touch the contacts or prongs on the AC plug within two seconds after unplugging the system from a wall socket. This precaution is to be adhered to until a Toshiba Service Representative replaces the power supply. Customers are to complete and return a Customer Reply Form and fax to 877-349-3054. Customers with questions regarding the notification should contact the Director of Regulatory Affairs at (800) 421-1968 or call their local Toshiba Representative at (800) 521-1968.
- Quantity
- 78
- Distribution
- Nationwide Distribution
- Lot, serial or model codes
- 4899 5131, 5132, 5133, 5134, 5224, 5139, 5260, 5261, 5369, 5256, 5367, 4883, 4884, 5370, 5374, 5375, 5264, 5257, 5135, 5223, 5379, 5262, 6283, 4886, 5258, 5382, 5383, 5384, 5385, 5216, 4887, 4897, 4898, 5027, 5380, 4982, 5026, 5137, 5138, 5219, 5220, 4979, 5465, 4900, 5466, 4978, 5377, 5136, 5376, 5072, 5071, 5221, 5215, 4981, 5371, 5372, 5373, 5029, 5030, 5068, 5069, 5070, 5222, 4980, 4885, 5259, 5028, 4896, 5368, 5217, 5218, 5140, 5265, 5364, 5365, 5366, 5381, 5467, 5468, 5469, 5470, 5471, 5472, 5473, 5474
- 510(k) numbers
- ["K091295"]
- PMA numbers
- not on file
- Product code
- IYO
- Firm FEI
- 2020563
- Firm city
- Tustin
- Firm state
- CA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28