Skip to the content

FDA Medical Device Recalls (openFDA)

Toshiba American Medical Systems Inc: Aplio XG model SSA-790A Aplio XG (SSA-790A) is a premium-class diagnostic ultrasound system incorporating state-of-the-art technologies. It is provided with the highest level of image quality and operability (reflecting the XG (next Generation) in its name), and can be used with ease for research as well as for routine clinical studies.

Recall number
Z-2981-2011
Recalling firm
Toshiba American Medical Systems Inc
Product
Aplio XG model SSA-790A Aplio XG (SSA-790A) is a premium-class diagnostic ultrasound system incorporating state-of-the-art technologies. It is provided with the highest level of image quality and operability (reflecting the XG (next Generation) in its name), and can be used with ease for research as well as for routine clinical studies.
Status
Terminated
Initiated
2010-06-02
Posted
2011-08-10
Terminated
2012-04-20
Reason
Toshiba America Medical Systems Inc (TAMS) is initiating a field correction on the the Aplio XG model SSA-790A because it did not meet the requirement of clause 15b, IEC60601-1 which requires that the residual voltage shall not exceed 60V after 1 second from taking off the AC plug.
Root cause
Device Design
Action
Toshiba America Medical Systems (TAMS) sent an "URGENT:MEDICAL DEVICE CORRECTION' letter dated June 2, 2010 to all affected customers. The letter describes the product, problem, and actions to be taken by the customers. Customers were instructed to be careful not to touch the contacts or prongs on the AC plug within two seconds after unplugging the system from a wall socket. This precaution is to be adhered to until a Toshiba Service Representative replaces the power supply. Customers are to complete and return a Customer Reply Form and fax to 877-349-3054. Customers with questions regarding the notification should contact the Director of Regulatory Affairs at (800) 421-1968 or call their local Toshiba Representative at (800) 521-1968.
Quantity
78
Distribution
Nationwide Distribution
Lot, serial or model codes
4899 5131, 5132, 5133, 5134, 5224, 5139, 5260, 5261, 5369, 5256, 5367, 4883, 4884, 5370, 5374, 5375, 5264, 5257, 5135, 5223, 5379, 5262, 6283, 4886, 5258, 5382, 5383, 5384, 5385, 5216, 4887, 4897, 4898, 5027, 5380, 4982, 5026, 5137, 5138, 5219, 5220, 4979, 5465, 4900, 5466, 4978, 5377, 5136, 5376, 5072, 5071, 5221, 5215, 4981, 5371, 5372, 5373, 5029, 5030, 5068, 5069, 5070, 5222, 4980, 4885, 5259, 5028, 4896, 5368, 5217, 5218, 5140, 5265, 5364, 5365, 5366, 5381, 5467, 5468, 5469, 5470, 5471, 5472, 5473, 5474
510(k) numbers
["K091295"]
PMA numbers
not on file
Product code
IYO
Firm FEI
2020563
Firm city
Tustin
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register