Skip to the content

FDA Medical Device Recalls (openFDA)

Medline Industries, Inc.: Cardiac Insulation Pad, OPEN HEART CDS

Recall number
Z-2979-2018
Recalling firm
Medline Industries, Inc.
Product
Cardiac Insulation Pad, OPEN HEART CDS
Status
Terminated
Initiated
2017-04-12
Posted
not on file
Terminated
2019-08-05
Reason
Complete delivery System (CDS) # CDS984746C was assembled using a non-sterile cardiac insulation pad, Medline component number 32369. The non-sterile component (32369) was placed within the non-sterile portion of the CDS assembly. This component should have been placed inside the sterile portion of the pack where it would have subsequently been exposed to an ethylene oxide sterilization process. This component was not placed in the sterile portion of the pack and was not sterilized, resulting in the inadvertent release of this item in a non-sterile state. The contents of the sterile kit were properly sterilized and the remaining contents of the non-sterile subassembly had the appropriate sterility status based on their intended use.
Root cause
Under Investigation by firm
Action
Firm contacted customers via phone and letter beginning April 12, 2017. Customers were instructed to remove the Ethicon BioPatch Protective Disk with CHG from the kit(s) and return it to the firm. The firm would provide a sterile replacement BioPatch. Affected product will be destroyed.
Quantity
77 kits
Distribution
CA, NM
Lot, serial or model codes
Model: CDS984745C; Lots: 17GD0936, 17FD1655, 17ED3419, 17ED1482, 17ED0636, 17DD0666, 17DD0296, 17AD1912, 16LD3005, 16LD0059, 16JD3201, 16ID3179, 16HD1298, 16GD0779, 16FD3925, 16CD0303 and 16AD4032
510(k) numbers
not on file
PMA numbers
not on file
Product code
N/A
Firm FEI
3000717925
Firm city
Waukegan
Firm state
IL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register