Skip to the content

FDA Medical Device Recalls (openFDA)

Medline Industries, Inc.: Laparoscopy CDS

Recall number
Z-2978-2018
Recalling firm
Medline Industries, Inc.
Product
Laparoscopy CDS
Status
Terminated
Initiated
2017-04-12
Posted
not on file
Terminated
2019-08-05
Reason
One manufacturing lot of a non-sterile pack was assembled with a non-sterile component labeled sterile (kit component #80532NS). The non-sterile component 80532NS was supposed to be placed inside the sterile portion of the pack where it would subsequently be exposed to the ethylene oxide sterilization process. However, the component was not placed in the sterile portion of the pack during assembly and was never exposed to the sterilization process. Component 80532NS was inadvertently released non-sterile. Component 80532NS is packaged by Medline and is an Anti-fog kit that contains 1 bottle of 6g Anti-Fog solution and an adhesive backed sponge.
Root cause
Under Investigation by firm
Action
Firm contacted customers via phone and letter beginning April 12, 2017. Customers were instructed to remove the Ethicon BioPatch Protective Disk with CHG from the kit(s) and return it to the firm. The firm would provide a sterile replacement BioPatch. Affected product will be destroyed.
Quantity
15 kits
Distribution
CA, NM
Lot, serial or model codes
Model: CDS920114F. Lot: 17MB5499
510(k) numbers
not on file
PMA numbers
not on file
Product code
N/A
Firm FEI
3000717925
Firm city
Waukegan
Firm state
IL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register