FDA Medical Device Recalls (openFDA)
Philips Healthcare Inc.: Allura Xper Systems Models: Allura Xper FD10 (F/C)-Biplane Fixed Allura Xper FD10/10 (incl. OR table) Allura Xper FD20- High resolution fixed-x ray Allura Xper FD20/10 Allura Xper FD20/20 Allura CV20 The Allura Xper F010, FD10/10, FD20, FD20/10 and FD20/20 Use From Labeling): is intended for use in cardiovascular and vascular X-ray imaging applications. including diagnostic,interventional procedures (such as PTCA, stent placement and atherectomies), pacemaker implantations and ElectroPhysiology (EP). The Allura Xper is intended for: Dedicated vascular and neurovascular imaging applications, including diagnostic and interventional procedures. This includes, e.g. peripheral, cerebral. thoracic and abdominal angiography, as well as PTCAs. stent placements, embolisations and thrombolysis. and Cardiac imaging applications including diagnostics, interventional procedures (such as PTCA, stent placing, atherectomies), pacemaker implantations and Electrophysiology (EP). Non-vascular interventions such as drainages, biopsies and vertebroplasties procedures. Use of the Allura Xper FD Series in the OR environment is restricted; the system should only be used for minimally invasive procedures. (Endovascular)
- Recall number
- Z-2976-2011
- Recalling firm
- Philips Healthcare Inc.
- Product
- Allura Xper Systems Models: Allura Xper FD10 (F/C)-Biplane Fixed Allura Xper FD10/10 (incl. OR table) Allura Xper FD20- High resolution fixed-x ray Allura Xper FD20/10 Allura Xper FD20/20 Allura CV20 The Allura Xper F010, FD10/10, FD20, FD20/10 and FD20/20 Use From Labeling): is intended for use in cardiovascular and vascular X-ray imaging applications. including diagnostic,interventional procedures (such as PTCA, stent placement and atherectomies), pacemaker implantations and ElectroPhysiology (EP). The Allura Xper is intended for: Dedicated vascular and neurovascular imaging applications, including diagnostic and interventional procedures. This includes, e.g. peripheral, cerebral. thoracic and abdominal angiography, as well as PTCAs. stent placements, embolisations and thrombolysis. and Cardiac imaging applications including diagnostics, interventional procedures (such as PTCA, stent placing, atherectomies), pacemaker implantations and Electrophysiology (EP). Non-vascular interventions such as drainages, biopsies and vertebroplasties procedures. Use of the Allura Xper FD Series in the OR environment is restricted; the system should only be used for minimally invasive procedures. (Endovascular)
- Status
- Terminated
- Initiated
- 2011-07-08
- Posted
- 2011-08-10
- Terminated
- 2014-05-14
- Reason
- System freezes from 20 seconds up to one minute. No fluoroscopy exposure and no geometry movements during this time period.
- Root cause
- Software design
- Action
- Philips Health Care North America sent a FIELD SAFETY CORRECTION ACTION letter dated July 4, 2011, to all affected customers. The letter identified the product, the problem and the action needed to be taken by the customer. This FCD will be identified as FCD72200200 and will be free of charge. This FCD consists a CD with the correct firmware and a installation procedure to implement the FCD correctly. Please contact your local Philips representative. For further follow up please call (978 ) 687-1501.
- Quantity
- 1423 units
- Distribution
- Nationwide Distribution
- Lot, serial or model codes
- Software releases: PBL 6.5.1 1PBL 6.5.2 1PBL 7.0.x 1PBL 7.2.x
- 510(k) numbers
- ["K031333","K033737","K041949"]
- PMA numbers
- not on file
- Product code
- IZI
- Firm FEI
- 1218950
- Firm city
- Andover
- Firm state
- MA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28