Skip to the content

FDA Medical Device Recalls (openFDA)

Philips Healthcare Inc.: Allura Xper Systems Models: Allura Xper FD10 (F/C)-Biplane Fixed Allura Xper FD10/10 (incl. OR table) Allura Xper FD20- High resolution fixed-x ray Allura Xper FD20/10 Allura Xper FD20/20 Allura CV20 The Allura Xper F010, FD10/10, FD20, FD20/10 and FD20/20 Use From Labeling): is intended for use in cardiovascular and vascular X-ray imaging applications. including diagnostic,interventional procedures (such as PTCA, stent placement and atherectomies), pacemaker implantations and ElectroPhysiology (EP). The Allura Xper is intended for: Dedicated vascular and neurovascular imaging applications, including diagnostic and interventional procedures. This includes, e.g. peripheral, cerebral. thoracic and abdominal angiography, as well as PTCAs. stent placements, embolisations and thrombolysis. and Cardiac imaging applications including diagnostics, interventional procedures (such as PTCA, stent placing, atherectomies), pacemaker implantations and Electrophysiology (EP). Non-vascular interventions such as drainages, biopsies and vertebroplasties procedures. Use of the Allura Xper FD Series in the OR environment is restricted; the system should only be used for minimally invasive procedures. (Endovascular)

Recall number
Z-2976-2011
Recalling firm
Philips Healthcare Inc.
Product
Allura Xper Systems Models: Allura Xper FD10 (F/C)-Biplane Fixed Allura Xper FD10/10 (incl. OR table) Allura Xper FD20- High resolution fixed-x ray Allura Xper FD20/10 Allura Xper FD20/20 Allura CV20 The Allura Xper F010, FD10/10, FD20, FD20/10 and FD20/20 Use From Labeling): is intended for use in cardiovascular and vascular X-ray imaging applications. including diagnostic,interventional procedures (such as PTCA, stent placement and atherectomies), pacemaker implantations and ElectroPhysiology (EP). The Allura Xper is intended for: Dedicated vascular and neurovascular imaging applications, including diagnostic and interventional procedures. This includes, e.g. peripheral, cerebral. thoracic and abdominal angiography, as well as PTCAs. stent placements, embolisations and thrombolysis. and Cardiac imaging applications including diagnostics, interventional procedures (such as PTCA, stent placing, atherectomies), pacemaker implantations and Electrophysiology (EP). Non-vascular interventions such as drainages, biopsies and vertebroplasties procedures. Use of the Allura Xper FD Series in the OR environment is restricted; the system should only be used for minimally invasive procedures. (Endovascular)
Status
Terminated
Initiated
2011-07-08
Posted
2011-08-10
Terminated
2014-05-14
Reason
System freezes from 20 seconds up to one minute. No fluoroscopy exposure and no geometry movements during this time period.
Root cause
Software design
Action
Philips Health Care North America sent a FIELD SAFETY CORRECTION ACTION letter dated July 4, 2011, to all affected customers. The letter identified the product, the problem and the action needed to be taken by the customer. This FCD will be identified as FCD72200200 and will be free of charge. This FCD consists a CD with the correct firmware and a installation procedure to implement the FCD correctly. Please contact your local Philips representative. For further follow up please call (978 ) 687-1501.
Quantity
1423 units
Distribution
Nationwide Distribution
Lot, serial or model codes
Software releases: PBL 6.5.1 1PBL 6.5.2 1PBL 7.0.x 1PBL 7.2.x
510(k) numbers
["K031333","K033737","K041949"]
PMA numbers
not on file
Product code
IZI
Firm FEI
1218950
Firm city
Andover
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register