Pro-Med Instruments Gmbh: DORO LUCENT Transitional Member, REF 1101.026 Component of headrest system used in open and percutaneous craniotomies and spinal surgeries

Recall number
Z-2972-2018
Recalling firm
Pro-Med Instruments Gmbh
Product
DORO LUCENT Transitional Member, REF 1101.026 Component of headrest system used in open and percutaneous craniotomies and spinal surgeries
Status
Terminated
Initiated
2018-07-05
Posted
2018-09-17
Terminated
2020-11-19
Reason
The combination of two different design revisions of DORO LUCENT Base Unit, Locking Transitional Member, and Transitional Member within one set up might lead to a malfunction and failure of the headrest system.
Root cause
Component change control
Action
The firm initiated the recall by email on 07/05/2018. This stage is only relevant for hospitals / users who are in possession with concerned products of two different revisions (mix-up): a) Discontinue the use of the combination and quarantine the concerned products b) Contact pro med instruments for further instructions relating to temporary replacement possibilities
Quantity
31 units
Distribution
Worldwide distribution to US states of CA, TX, TN, MS, GA, WV, and OH, Qatar, India, UK, Singapore, Australia, Germany, South Africa, Italy, and Japan
Lot, serial or model codes
Serial Numbers: 1704002 1704004 1704005 1704007 1704011 1704041 1704010 1704019 1704012 1704014 1704016 1704017 1704021 1704024 1704025 1704036 1704038 1704018 1704008 1704031 1704026 1704027 1704028 1704029 1704030 1704032 1704034 1704035 1704006 1704037 1704039
510(k) numbers
["K032331"]
PMA numbers
not on file
Product code
HBL
Firm FEI
3003923584
Firm city
Freiburg Im Breisgau
Firm state
not on file
Firm country
Germany

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register