Skip to the content

FDA Medical Device Recalls (openFDA)

AMO Manufacturing USA, LLC: STAR S4 IR Excimer Laser System; Software Version 5.18 Mfg by : AMO Manufacturing USA, LLC 510 Cottonwood Drive Milpitas, CA 95035 Iris registration is designed to compensate for pupil centroid shift and cycloration, both of Which may occur because of patient characteristics and the different environments at the time of measurement and at the time of treatment.

Recall number
Z-2971-2011
Recalling firm
AMO Manufacturing USA, LLC
Product
STAR S4 IR Excimer Laser System; Software Version 5.18 Mfg by : AMO Manufacturing USA, LLC 510 Cottonwood Drive Milpitas, CA 95035 Iris registration is designed to compensate for pupil centroid shift and cycloration, both of Which may occur because of patient characteristics and the different environments at the time of measurement and at the time of treatment.
Status
Terminated
Initiated
2011-07-18
Posted
2011-08-09
Terminated
2013-09-10
Reason
Reports of lower success rates with Iris Registration (IR) image capture with the STAR S4 IR Excimer Laser System v 5.3 that with previous version.
Root cause
Software design
Action
Abbott Medical Optics (AMO) sent an "ADVISORY NOTICE" dated July 18, 2011 to all customers using the AMO WAVESCAN Wavefront System. The letter identifies the product, problem, and actions to be taken by the customers. The letter instructs customers to create a separate patient file for each eye on the WAVESCAN System (STAR S4 IR Excimer Laser Systems,software versions 5.00-5.30) according to the directions in attachment A. Additionally, customers are to complete and return an attached form to AMO via fax at the regional fax number. Contact an AMO representative for questions regarding this notice.
Quantity
673 total - v5.00 -5.22
Distribution
Worldwide Distribution-USA (nationwide) including Puerto Rico and countries of Africa, Asian Pacific, Canada, Europe, Japan, Latin America, and the Middle East.
Lot, serial or model codes
Model # 0030-4817
510(k) numbers
not on file
PMA numbers
["P930016S019"]
Product code
LZS
Firm FEI
3006695864
Firm city
Milpitas
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register