Skip to the content

FDA Medical Device Recalls (openFDA)

Paragonix Technologies Inc.: Paragonix Technologies Inc. KidneyVault Portable Renal Perfusion System REF# PRGNX-8008-001

Recall number
Z-2967-2026
Recalling firm
Paragonix Technologies Inc.
Product
Paragonix Technologies Inc. KidneyVault Portable Renal Perfusion System REF# PRGNX-8008-001
Status
Open, Classified
Initiated
2026-07-20
Posted
2026-08-14
Terminated
not on file
Reason
Due to a firmware defect, the portable organ transfer device datalogger may lose its configuration, resulting in a blank screen and loss of displayed preservation parameters.
Root cause
Software design
Action
On July 20, 2026, Paragonix Technologies Inc./Getinge issued an Urgent Medical Device Correction to affected consignees via FedEX mail. Paragonix requested consignees to take the following actions: 1. An update to the Paragonix App has been released through the application store. 2. We recommend updating all mobile devices to latest version prior to KidneyVault usage. 3. Please continue to inspect all KidneyVault devices per the IFU before use to ensure proper datalogger function. 4. Continue to follow the IFU for using a multi-faceted clinical approach to assessing the organ, including any/all information available on the datalogger screen. 5. Please ensure this message is forwarded to any individuals that need notification within your organization or any organization where the affected devices have been transferred. 6. Complete and return the enclosed response form by e-mailing a copy to recallresponses.qrc@getinge.com.
Quantity
1445 systems
Distribution
US Nationwide distribution in the states and territories of AL, AZ, CA, TN, FL, HI, MN, NJ, TX, UT, CO, CT, DC, MD, GA, KY, LA, MA, MI, MO, MS, NC, ND, NE, NM, NV, NY, PA, PR, SC, UT, VA, WI, WV, IN, OH, KY, IL, MD, MN, OR.
Lot, serial or model codes
All Serial Numbers/UDI: 00850005470342 Software Version: Paragonix App Ver. 6.1.1/6.1.2
510(k) numbers
["K234060"]
PMA numbers
not on file
Product code
KDN
Firm FEI
3011175555
Firm city
Braintree
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register