FDA Medical Device Recalls (openFDA)
Siemens Healthcare Diagnostics Products Ltd.: Siemens Healthineers Immulite 2000 Systems SYP Syphilis Screen REF(s):L2kSY2/10489085; L2KSY6/10488319 IVD, RxOnly
- Recall number
- Z-2966-2026
- Recalling firm
- Siemens Healthcare Diagnostics Products Ltd.
- Product
- Siemens Healthineers Immulite 2000 Systems SYP Syphilis Screen REF(s):L2kSY2/10489085; L2KSY6/10488319 IVD, RxOnly
- Status
- Open, Classified
- Initiated
- 2026-07-13
- Posted
- 2026-08-14
- Terminated
- not on file
- Reason
- Syphilis Screen assay may contain a negative patient bias resulting in low reactive patient results being reported as nonreactive.
- Root cause
- Under Investigation by firm
- Action
- On July 13, 2026, Siemens Healthineers issued a Urgent Medical Device Correction to affected consignees via FedEx. Siemens asked consignees to take the following actions: 1. Review this letter with your Medical Director to determine the appropriate course of action, including for any previously generated results, if applicable. 2.Discontinue use and discard the affected kit lot. 3. Review your current inventory to determine your laboratory s replacement needs and provide this information to Siemens Healthineers for reporting to the authorities. 4. Complete and return the Field Correction Effectiveness Check Form attached to this letter within (30) days. 5.Retain this letter with your laboratory records and forward this letter to those who may have received the affected product(s).
- Quantity
- 478 units
- Distribution
- Worldwide - US Nationwide distribution in the states of FL, NJ, NY and the countries of Austria, Azerbaijan, Belgium, Brazil, Chile, Colombia, Cura-ao, St.¿Eus Cyprus, France, Germany, Italy, Kazakhstan, Latvia, Lithuania, Netherlands, Romania, Russian¿Fed., Saudi¿Arabia, Slovakia, Slovenia, South¿Africa, Spain, U.A.E., Ukraine, United¿Kingdom, Uzbekistan.¿¿¿¿¿
- Lot, serial or model codes
- Lot Code: D387/UDI(s):L2KSY2-00630414999791; L2KSY6 - 00630414999807
- 510(k) numbers
- ["K091361"]
- PMA numbers
- not on file
- Product code
- LIP
- Firm FEI
- 3002806944
- Firm city
- Caernarfon
- Firm state
- not on file
- Firm country
- United Kingdom
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28