Haag-Streit USA Inc: Lenstar LS 900 EyeSuite Biometry version i2.000, Four (4) Protective Cardboard Containers Per Each System. Haag-Streit USA, Inc. A non-invasive, non-contact OLCR (Optical Low Coherence Reflectometer) Biometer used for obtaining ocular measurements and performing calculations to assist in determination of the appropriate power and type of IOL (Intraocular Lens) for implantation after removal of the natural crystalline lens following cataract removal.

Recall number
Z-2964-2011
Recalling firm
Haag-Streit USA Inc
Product
Lenstar LS 900 EyeSuite Biometry version i2.000, Four (4) Protective Cardboard Containers Per Each System. Haag-Streit USA, Inc. A non-invasive, non-contact OLCR (Optical Low Coherence Reflectometer) Biometer used for obtaining ocular measurements and performing calculations to assist in determination of the appropriate power and type of IOL (Intraocular Lens) for implantation after removal of the natural crystalline lens following cataract removal.
Status
Terminated
Initiated
2011-06-20
Posted
2011-08-09
Terminated
2013-03-08
Reason
A patient record may be linked to an incorrect/inconsistent set of calculation results under specific circumstances when using EyeSuite Software version i2.000.
Root cause
Software change control
Action
Haag-Streit USA, Inc. sent an "Urgent Field Safety Notices" dated June 20, 2011 to all affected customers. The letter included a description of the affected product and problem. The firm instructed their customers to make sure all new and current templates are saved and upgrade to EyeSuite version i4.000. Customers are asked to complete and return the Software Update Verification Form. For additional information contact the Haag-Streit at (513) 398-3937.
Quantity
216
Distribution
Nationwide (USA) Distribution.
Lot, serial or model codes
Part Number 7220162.  Serial Numbers: 21, 127, 223, 224, 284, 285, 316, 318, 319, 320, 322, 323, 324, 325, 326, 327, 328, 330, 331, 332, 333, 364, 365, 366, 368, 369, 371, 372, 376, 378, 379, 381, 384, 385, 386, 405, 407, 408, 409, 411, 412, 413, 414, 415, 416, 417, 418, 419, 420, 421, 423, 424, 443, 446, 447, 448, 449, 450, 452, 453, 454, 455, 456, 457, 458, 459, 460, 463, 469, 470, 481, 482, 483, 484, 485, 486, 487, 489, 490, 491, 492, 493, 494, 495, 496, 497, 499, 500, 527, 528, 529, 531, 532, 533, 535, 536, 537, 538, 539, 540, 541, 542, 543, 544, 545, 546, 547, 548, 549, 550, 551, 552, 558, 559, 560, 561, 563, 564, 565, 566, 567, 568, 569, 570, 571, 572, 573, 574, 576, 578, 579 581, 582, 583, 584, 585, 587, 588, 589, 590, 618, 619, 620, 621, 622, 624, 625, 626, 628, 630, 631, 632, 633, 634, 635, 636, 638, 639, 640, 641, 643, 644, 645, 647, 648, 649, 650, 651, 652, 653, 654, 656, 657, 658, 659, 660, 661, 662, 663, 665, 667, 668, 669, 670, 671, 672, 674, 675, 676, 677, 721, 723, 724, 725, 726, 727, 728, 729, 735, 824, 825, 922, 946, 966, 967, 968, 970, 1018, 1019, 1021, 1022, 1024, 1025, 1114, 1143 & 1195.
510(k) numbers
["K082891"]
PMA numbers
not on file
Product code
HJB
Firm FEI
1000136533
Firm city
Mason
Firm state
OH
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register