Philips North America Llc: Cardiac Workstation 5000. Product Number: 860439.
- Recall number
- Z-2955-2026
- Recalling firm
- Philips North America Llc
- Product
- Cardiac Workstation 5000. Product Number: 860439.
- Status
- Open, Classified
- Initiated
- 2026-07-01
- Posted
- 2026-08-14
- Terminated
- not on file
- Reason
- Device will intermittently provide inaccurate data related to native heart rate.
- Root cause
- Software design
- Action
- Philips notified consignees on about 07/01/2026 via URGENT Medical Device Correction letter. Consignees were informed that they could continue the use of the device but to refer to the information within the notification letter, disable the automatic Standby feature on the Cardiac Workstation and avoid manually putting the Cardiac Workstation in Standby mode to minimize the inaccurate heart rate value from occurring, and to pass the notice to all those who need to be aware within each organization or to any organization where affected product(s) have been potentially transferred. Additionally, consignees were requested to place the Urgent Medical Device Correction letter with the documentation of the Cardiac Workstation and associated devices in a place where it is most likely to be seen and viewed. Philips representatives will contact consignees to arrange an upgrade to the Cardiac Workstation Release 1.2.0.7.
- Quantity
- 796 units
- Distribution
- Worldwide distribution - US Nationwide and the countries of Australia, Austria, Bahrain, Belgium, Brazil, Canada, Croatia, Czech Republic, Dominican Republic, France, Germany, Gibraltar, Hong Kong, Iceland, Ireland, Iraq, Italy, Luxembourg, Malta, Netherlands, New Caledonia, New Zealand, Norway, Poland, Portugal, Qatar, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, United Arab Emirates, United Kingdom.
- Lot, serial or model codes
- Product Number: 860439. UDI: (01)00884838094826. Serial Numbers: US42510185, US42510184, US52510209, US52510200, US52510226, US52510228, US62510247, US62510250, US62510256, US72510286, US82510299, US82510307, US82510300, US82510313, US82510318, US82510314, US82510335, US82510336, US82510341, US82510325, US92510351, US92510353, US92510354, US92510361, US92510362, US92510370, USN2510413, USN2510414, USN2510434, USN2510435, USN2510437, USD2510506, USD2510519, US12610582, USD2510513, US12610598, US22610609, US22610683, US22610691, US22610689, US32610737, US32610744, US32610759, US32610766, US32610749, US32610739, US32610751, USD2510516, USD2510510, USD2510517, USD2510529, USD2510509, US32610752, US32610753, US32610879, US32610880, US32610901, US32610914, US42610946, US32610915, US32610941, US32610932, US42610943, US42610942, US42610959, US42610958, US42610955, US42610974, US42610967, US42610956, US42610993, US42610960, CN72410032, CN92410061, CNO2410064, CNO2410068, CNN2410075, CND2410080, CND2410101, CND2410103, CN12510116, CN12510136, CN22510140, CN22510142, CN22510146, CN32510148, CN32510161, CN42510168, CN42510181, CN42510183, CN52510204, CN52510201, CN52510217, CN62510239, CN62510241, CN62510242, CN62510253, CN62510254, CN62510251, CN72510275, CN72510282, CN72510281, CN72510285, CN72510289, CN72510288, CN72510293, CN72510297, CN82510305, CN82510310, CN82510323, CN82510324, CN82510321, CN92510343, CN92510364, CN92510365, CN92510367, CN92510374, CNO2510380, CNO2510392, CNO2510383, CNO2510386, CNO2510396, CNN2510412, CNN2510410, CNN2510427, CNN2510443, CNN2510444, CNN2510454, CNN2510485, CNN2510486, CND2510503, CND2510502, CND2510508, CND2510528, CND2510530, CND2510566, CND2510575, CND2510574, CN12610585, CN12610590, CN12610595, CN12610593, CN22610612, CN22610681, CN22610707, CN22610735, CN32610755, CN32610760, CN32610761, CN32610762, CN32610763, CN32610765, CN32610818, CN32610826, CN32610910, CN32610913, CN32610912, CN32610924, CN32610923, CN32610929, CN32610930, CN32610933, CN42610944, CN42610951, CN42610970, US42610994
- 510(k) numbers
- ["K241556"]
- PMA numbers
- not on file
- Product code
- DPS
- Firm FEI
- 3016618143
- Firm city
- Cambridge
- Firm state
- MA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28