FDA Medical Device Recalls (openFDA)
Integra LifeSciences Corp.: Integra Padgett Electric Dermatome Set - Product Usage: are intended for use in the cutting of skin grafts for use in plastic surgery and/or burn treatment. These instruments are intended only for professional use in hospitals and similar medical facilities, where the patient will be under the supervision of trained personnel.
- Recall number
- Z-2955-2020
- Recalling firm
- Integra LifeSciences Corp.
- Product
- Integra Padgett Electric Dermatome Set - Product Usage: are intended for use in the cutting of skin grafts for use in plastic surgery and/or burn treatment. These instruments are intended only for professional use in hospitals and similar medical facilities, where the patient will be under the supervision of trained personnel.
- Status
- Terminated
- Initiated
- 2020-08-04
- Posted
- 2020-09-12
- Terminated
- 2021-08-12
- Reason
- lntegra LifeSciences has identified through a number of complaints that the handpiece cable melts following sterilization.
- Root cause
- Mixed-up of materials/components
- Action
- All impacted consignees will be notified by letter delivered by courier service, facsimile or email about the initial notification between August 4th and 6th, 2020. Customers are asked to remove the affected product from use and complete and return the response form to the manufacturer. After response forms are reviewed the customer will be asked to return the product and a replacement will be sent.
- Quantity
- 2 units
- Distribution
- Worldwide distribution - US Nationwide distribution including in the states of MN, TX, CA, UT, WV, FL, IN, GA, AZ, and the countries of Latin America ( CR, MX, TT, EC, PA ), Canada, Europe (FR, AT, BE, ES).
- Lot, serial or model codes
- Catalog Number: 3539900 Catalog Description: Electric Dermatome Set Model #: S6 Lot/Serial #: S-594-6 NOT SOLD IN USA
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- GFD
- Firm FEI
- 3003418325
- Firm city
- Princeton
- Firm state
- NJ
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as