Skip to the content

FDA Medical Device Recalls (openFDA)

Foundation Medicine, Inc.: Hamilton Microlab STAR Liquid Handler; Protocol Configuration Tool (PCT) v2.0.0. Used to conduct F1CDx and F1LCDx.

Recall number
Z-2951-2026
Recalling firm
Foundation Medicine, Inc.
Product
Hamilton Microlab STAR Liquid Handler; Protocol Configuration Tool (PCT) v2.0.0. Used to conduct F1CDx and F1LCDx.
Status
Open, Classified
Initiated
2026-06-10
Posted
2026-08-14
Terminated
not on file
Reason
Software issue that can cause an increase in spontaneous software closures.
Root cause
Software design
Action
An URGENT MEDICAL DEVICE RECALL NOTIFICATION dated 6/10/26 was emailed to consignees. Consignees were instructed to review the impacted equipment list and verify that the PCT V2.0.1 software version has been installed. If software hasn't been updated FMI will follow up to assist with installation. The recall notice is to be forwarded to anyone in the facility or outside the organization that needs to be informed. The notification is to be retained with laboratory records. Questions can be directed to Foundation Medicine at 888-988-3639 or QACTAdverseEventReporting@foundationmedicine.com.
Quantity
32 units
Distribution
US Nationwide distribution in the states of MA & NC.
Lot, serial or model codes
Software Version: Protocol Configuration Tool (PCT) v2.0.0; Equipment ID: 100071, 100213, 100355, 102281, 103047, 104154, 104201, 104202, 104204, 104205, 104224, 104503, 104504, 104562, 104885, 105961, 106497, 00016, 00487, 00488, 00489, 00490, 00817, 00491, 00492, 00854, 01173, 01174, 01785, 01337, 01596, 01993.
510(k) numbers
not on file
PMA numbers
["P200006","P170019"]
Product code
PQP
Firm FEI
3010679023
Firm city
Cambridge
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register