FDA Medical Device Recalls (openFDA)
Foundation Medicine, Inc.: Hamilton Microlab STAR Liquid Handler; Protocol Configuration Tool (PCT) v2.0.0. Used to conduct F1CDx and F1LCDx.
- Recall number
- Z-2951-2026
- Recalling firm
- Foundation Medicine, Inc.
- Product
- Hamilton Microlab STAR Liquid Handler; Protocol Configuration Tool (PCT) v2.0.0. Used to conduct F1CDx and F1LCDx.
- Status
- Open, Classified
- Initiated
- 2026-06-10
- Posted
- 2026-08-14
- Terminated
- not on file
- Reason
- Software issue that can cause an increase in spontaneous software closures.
- Root cause
- Software design
- Action
- An URGENT MEDICAL DEVICE RECALL NOTIFICATION dated 6/10/26 was emailed to consignees. Consignees were instructed to review the impacted equipment list and verify that the PCT V2.0.1 software version has been installed. If software hasn't been updated FMI will follow up to assist with installation. The recall notice is to be forwarded to anyone in the facility or outside the organization that needs to be informed. The notification is to be retained with laboratory records. Questions can be directed to Foundation Medicine at 888-988-3639 or QACTAdverseEventReporting@foundationmedicine.com.
- Quantity
- 32 units
- Distribution
- US Nationwide distribution in the states of MA & NC.
- Lot, serial or model codes
- Software Version: Protocol Configuration Tool (PCT) v2.0.0; Equipment ID: 100071, 100213, 100355, 102281, 103047, 104154, 104201, 104202, 104204, 104205, 104224, 104503, 104504, 104562, 104885, 105961, 106497, 00016, 00487, 00488, 00489, 00490, 00817, 00491, 00492, 00854, 01173, 01174, 01785, 01337, 01596, 01993.
- 510(k) numbers
- not on file
- PMA numbers
- ["P200006","P170019"]
- Product code
- PQP
- Firm FEI
- 3010679023
- Firm city
- Cambridge
- Firm state
- MA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as