FDA Medical Device Recalls (openFDA)
Haemonetics Corporation: Haemonetics OrthoPAT System-OPAT QuickConnect Reservoirs. Product Code: OPT-R-1000 The OrthoPAT orthopedic perioperative autotransfusion system indicated for use to salvage RBCs from blood lost intraoperatively and postoperatively during surgical procedures
- Recall number
- Z-2951-2018
- Recalling firm
- Haemonetics Corporation
- Product
- Haemonetics OrthoPAT System-OPAT QuickConnect Reservoirs. Product Code: OPT-R-1000 The OrthoPAT orthopedic perioperative autotransfusion system indicated for use to salvage RBCs from blood lost intraoperatively and postoperatively during surgical procedures
- Status
- Terminated
- Initiated
- 2018-07-13
- Posted
- not on file
- Terminated
- 2020-06-16
- Reason
- OrthoPAT reservoirs are missing the cannula inside the reservoirs. The absence of this cannula prevents the emptying of blood from the reservoirs. The inability to empty the reservoir means that the collected blood cannot be processed.
- Root cause
- Process control
- Action
- The firm, Haemonetics, notified customers by an email titled "URGENT: MEDICAL DEVICE RECALL" on July 13, 2018 . The email described the product, problem, and actions to be taken. The customers were instructed to inspect product for defect and return units of the affected lots for replacement with non-affected units; complete the attached customer acknowledgement form via FAX TO +1-781-356-3558 OR SCAN AND E-MAIL TO: CORPORATEREGULATORY@HAEMONETICS.COM. You can contact your local customer service representative at: North America 1.800.537.2802 or email:customerservicena@haemonetics.com.
- Quantity
- 348 units
- Distribution
- US Nationwide Distribution
- Lot, serial or model codes
- Lot # 1117066
- 510(k) numbers
- ["K122262"]
- PMA numbers
- not on file
- Product code
- CAC
- Firm FEI
- 1219343
- Firm city
- Braintree
- Firm state
- MA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28