FDA Medical Device Recalls (openFDA)
Erbe USA Inc: ERBECRYO 2 Cryosurgical Unit Model/Catalog Number: 10402-000
- Recall number
- Z-2938-2026
- Recalling firm
- Erbe USA Inc
- Product
- ERBECRYO 2 Cryosurgical Unit Model/Catalog Number: 10402-000
- Status
- Open, Classified
- Initiated
- 2026-06-26
- Posted
- 2026-08-12
- Terminated
- not on file
- Reason
- A strong pull on the high-pressure transfer hose of cryosurgical unit may lead to pneumatic short circuit between he gas inlet and the connector resulting in CO2 leaking into the return path, leading to excessive pressure causing the cryoprobe to rupture.
- Root cause
- Device Design
- Action
- On June 26, 2026, Erbe USA issued a Urgent: Medical Device Correction update notification to affected consignees via multiple methods ( Email/Website/Hand delivery). ERBE ask consignees to take the following actions: 1. We request the ERBECRYO 2 Unit be made available to the Erbe Representative to install the (V)1.0.4 Software. The Erbe Representative will perform all tasks required to update the software. 2. Confirm acknowledgement with Erbe Representative on-site. 3. Share this notice with all users within your facility. 4.While awaiting software update, there is no change of the Cryosurgical Unit s intended use or performance. Please continue to follow all warnings, instructions, guidelines, etc. involved with the Unit and Cryoprobe.
- Quantity
- 11226 units
- Distribution
- US Nationwide distribution including in the states and territories of Alabama, Alaska, Arizona, Arkansas, California, Colorado, Connecticut, Delaware, District of Columbia, Florida, Georgia, Hawaii, Idaho, Illinois, Indiana, Iowa, Kansas, Kentucky, Louisiana, Maine, Maryland, Massachusetts, Michigan, Minnesota, Mississippi, Missouri, Montana, Nebraska, Nevada, New Hampshire, New Jersey, New Mexico, New York, North Carolina, North Dakota, Ohio, Oklahoma, Oregon, Pennsylvania, Puerto Rico, Rhode Island, South Carolina, South Dakota, Tennessee, Texas, Utah, Vermont, Virginia, Washington, West Virginia, Wisconsin, Wyoming.
- Lot, serial or model codes
- UDI-DI 04050147013797/ Serial Numbers: All serial numbers up to 11608268 Software Version: 1.0.3
- 510(k) numbers
- ["K190651"]
- PMA numbers
- not on file
- Product code
- GEH
- Firm FEI
- 1000140388
- Firm city
- Marietta
- Firm state
- GA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28