Skip to the content

FDA Medical Device Recalls (openFDA)

Erbe USA Inc: ERBECRYO 2 Cryosurgical Unit Model/Catalog Number: 10402-000

Recall number
Z-2938-2026
Recalling firm
Erbe USA Inc
Product
ERBECRYO 2 Cryosurgical Unit Model/Catalog Number: 10402-000
Status
Open, Classified
Initiated
2026-06-26
Posted
2026-08-12
Terminated
not on file
Reason
A strong pull on the high-pressure transfer hose of cryosurgical unit may lead to pneumatic short circuit between he gas inlet and the connector resulting in CO2 leaking into the return path, leading to excessive pressure causing the cryoprobe to rupture.
Root cause
Device Design
Action
On June 26, 2026, Erbe USA issued a Urgent: Medical Device Correction update notification to affected consignees via multiple methods ( Email/Website/Hand delivery). ERBE ask consignees to take the following actions: 1. We request the ERBECRYO 2 Unit be made available to the Erbe Representative to install the (V)1.0.4 Software. The Erbe Representative will perform all tasks required to update the software. 2. Confirm acknowledgement with Erbe Representative on-site. 3. Share this notice with all users within your facility. 4.While awaiting software update, there is no change of the Cryosurgical Unit s intended use or performance. Please continue to follow all warnings, instructions, guidelines, etc. involved with the Unit and Cryoprobe.
Quantity
11226 units
Distribution
US Nationwide distribution including in the states and territories of Alabama, Alaska, Arizona, Arkansas, California, Colorado, Connecticut, Delaware, District of Columbia, Florida, Georgia, Hawaii, Idaho, Illinois, Indiana, Iowa, Kansas, Kentucky, Louisiana, Maine, Maryland, Massachusetts, Michigan, Minnesota, Mississippi, Missouri, Montana, Nebraska, Nevada, New Hampshire, New Jersey, New Mexico, New York, North Carolina, North Dakota, Ohio, Oklahoma, Oregon, Pennsylvania, Puerto Rico, Rhode Island, South Carolina, South Dakota, Tennessee, Texas, Utah, Vermont, Virginia, Washington, West Virginia, Wisconsin, Wyoming.
Lot, serial or model codes
UDI-DI 04050147013797/ Serial Numbers: All serial numbers up to 11608268 Software Version: 1.0.3
510(k) numbers
["K190651"]
PMA numbers
not on file
Product code
GEH
Firm FEI
1000140388
Firm city
Marietta
Firm state
GA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register