FDA Medical Device Recalls (openFDA)
Covidien, LP: Dermalon 88861757-31 DERMALON* 4-0 BLU 75CM C13 88861757-31 88861799-31 DERMALON* 4-0 BLU 75CM C14 88861799-31
- Recall number
- Z-2920-2026
- Recalling firm
- Covidien, LP
- Product
- Dermalon 88861757-31 DERMALON* 4-0 BLU 75CM C13 88861757-31 88861799-31 DERMALON* 4-0 BLU 75CM C14 88861799-31
- Status
- Open, Classified
- Initiated
- 2026-06-24
- Posted
- 2026-08-07
- Terminated
- not on file
- Reason
- Product may contain a perforation or tear in the sterile barrier packaging. This issue may result in biocontamination of the sterile sutures, which may lead to infection, loss of vision if in ophthalmic utilization, or prolonged surgery.
- Root cause
- Equipment maintenance
- Action
- The firm began notifying consignees on 7/7/26 via email, mail, or electronically. Notices were titled, "URGENT: VOLUNTARY MEDICAL DEVICE RECALL." Customers were informed of the manufacturing nonconformance. If products in scope of this field action have been used in patients under the customer's care, the customer is instructed to monitor for signs of infection or vision issues due to the possibility of biocontamination. Customers were instructed to immediately identify, cease use, and quarantine all unused impacted product. Unused impacted product is to be returned to Medtronic. The recall notice should be passed on to all those who need to be aware within the customer's organization or to any organization where devices have been transferred.
- Quantity
- 135,875 total
- Distribution
- US Nationwide and the countries of Australia, Austria, Belgium, Bulgaria, Canada, Canary Islands, Chile, China, Colombia, Croatia, Cyprus, Czech Republic, Denmark, Estonia, Finland, France, Germany, Greece, Guadeloupe, Guatemala, Honduras, Hong Kong, Hungary, Iceland, India, Ireland, Israel, Italy, Japan, Kosovo, Lebanon, Lithuania, Luxembourg, Mexico, Namibia, Netherlands, North Macedonia, Northern Ireland, Norway, Oman, Panama, Poland, Portugal, Qatar, Reunion, Romania, Russia, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, Ukraine, United Arab Emirates, United Kingdom, Uzbekistan.
- Lot, serial or model codes
- 88861757-31 UDI-DI 10884521070912 Lot D6A3020FY UDI-DI 20884521070919 Lot D6A3020FY 88861799-31 UDI-DI 20884521071060 Lot D6A2861FY
- 510(k) numbers
- ["K981582"]
- PMA numbers
- not on file
- Product code
- GAR
- Firm FEI
- 1219930
- Firm city
- North Haven
- Firm state
- CT
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28