Skip to the content

FDA Medical Device Recalls (openFDA)

CareFusion 303, Inc.: PC Unit Front Case with Keypad Replacement Kits: TC10008389 ASSY CASE FRONT W/KEYPAD 8015LS, TC10010217 ASSY FRT CASE W/ KEYPAD 8015 M2, TC10012515 ASSY FR CASE W/ KEYPAD 8015 M2, TC10013702 ASSY, CASE, FRONT W/KEYPAD, 8015LS, TC10013664 ASSY FR CASE W/ KEYPAD 8015 M2

Recall number
Z-2910-2020
Recalling firm
CareFusion 303, Inc.
Product
PC Unit Front Case with Keypad Replacement Kits: TC10008389 ASSY CASE FRONT W/KEYPAD 8015LS, TC10010217 ASSY FRT CASE W/ KEYPAD 8015 M2, TC10012515 ASSY FR CASE W/ KEYPAD 8015 M2, TC10013702 ASSY, CASE, FRONT W/KEYPAD, 8015LS, TC10013664 ASSY FR CASE W/ KEYPAD 8015 M2
Status
Open, Classified
Initiated
2020-08-04
Posted
2020-09-04
Terminated
not on file
Reason
The pump module keypad may exhibit keys that are unresponsive or stuck as a result of fluid ingress potentially resulting in a delay to the start of infusion, interruption of infusion, or inability to titrate medication. For high-risk patient populations who are receiving life-sustaining infusions, delays in an infusion or interruption can cause serious injury or death.
Root cause
Reprocessing Controls
Action
On 08/04/20, Urgent Medical Device Recall notices were mailed to Clinical Users: If the PC unit keypad becomes unresponsive or a key gets stuck, clinicians should remove the device from service and send to Biomedical Engineering. If administering a critical medication, continue the infusion until it is safe to replace the PC unit. Cleaning personnel can minimize the potential for fluid ingress during cleaning by 1) not using a cloth that drips, 2) wringing out the cleaning cloth to squeeze out excess fluid, and 3) not spraying fluids directly onto the device. Biomedical Engineering: If an affected PC unit has an unresponsive keypad or stuck key, remove the unit from service and call the recalling firm for information replacement parts. Until 510(k) clearance, the affected PC unit keypad design will continue to be provided for replacement or repair. Inspect all PC unit keypads for bubbling, cracks, damage, layer separations, and holes. Replace the front case keypad if it fails inspection. Return the completed Customer Response Form. Provide a copy of this letter and enclosures to customers who are currently renting affected devices. Firm is currently working on a design change for the PC unit keypad and will notify customers when replacement keypads are available. Additional Information: Customer Advocacy, 888-812-3266, customerfeedback@bd.com; Training Resources www.bd.com/alaris-system-hardware-recall; Recall Support Center, 888-562-6018, SupportCenter@bd.com; Technical Support, 888-812-3229, DL-US-INF-TechSupport@bd.com; Customer Order Management, 1-800-482-4822, GMB-CTS-CustCareInfusion@bd.com On 12/16/22, Medical Device Notifications were sent to customers informing them that they may have received impacted parts for repair purposes or received product with impacted parts. Customers were provided with links to the original recall notices and asked to review the information and complete the customer response form.
Quantity
28346
Distribution
U.S.:LA, FL, KY, MI, AL, GA, CA, TN, NC, TX, MA, OH, CO, WI, IL, KS, MO, UT, AK, MD, NY, NV, VA, NE, IN, PA, VT, DC, SC, CT, AR, NJ, MT, ND ,HI , NH, WA, DE, ME, OR, MN, AZ, MS, IA, NM, WY, ID, OK, SD, WV, MP, GU O.U.S.: CA, AE, AU, BE, CA, CH, IL, IN, KW, N/A, NL, NZ, PH, QA, SA, SG, TW, US, ZA
Lot, serial or model codes
All kits manufactured from April 07, 2017 to Present.
510(k) numbers
["K133532"]
PMA numbers
not on file
Product code
FRN
Firm FEI
2016493
Firm city
San Diego
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register