Skip to the content

FDA Medical Device Recalls (openFDA)

Integra LifeSciences Corp.: CUSA Dissectron Fingertip Laparoscopic Cannula, REF F75221, Sterile EO, Rx only, packaged in a Tyvek pouch and secondary packaged in Tyvek pouch, Five (5) sealed pouches are packaged in a cardboard box. For use with CUSA DISSECTRON Hand pieces and Sontrodes

Recall number
Z-2906-2016
Recalling firm
Integra LifeSciences Corp.
Product
CUSA Dissectron Fingertip Laparoscopic Cannula, REF F75221, Sterile EO, Rx only, packaged in a Tyvek pouch and secondary packaged in Tyvek pouch, Five (5) sealed pouches are packaged in a cardboard box. For use with CUSA DISSECTRON Hand pieces and Sontrodes
Status
Terminated
Initiated
2016-08-23
Posted
not on file
Terminated
2017-03-29
Reason
During final sampling inspection, one of the lots was discovered with a total incomplete packaging seal of the sterile product.
Root cause
Process control
Action
Integra sent an "Urgent Voluntary Medical Device Recall" Letter and Recall Acknowledgement and Return Form dated August 23, 2016 to their affected customer. The letter identified the product, the problem, and the action to be taken by the customer. The customer was advised if they have distributed the product to identify those customers and forward a copy of the enclosed Integra Medical Device Voluntary Recall Notification to those customers. The customer was also asked to review their inventory for the affected product and complete the attached form indicating whether or not they have any of the affected product in their inventory. Customers with questions were instructed to contact Customer Service at 1-800-654-2873.
Quantity
9
Distribution
Nationwide Distribution to KS only
Lot, serial or model codes
Catalog Number F75221, Lot #0195486 with expiry 04/2021
510(k) numbers
not on file
PMA numbers
not on file
Product code
JXG
Firm FEI
3003418325
Firm city
Plainsboro
Firm state
NJ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register