Skip to the content

FDA Medical Device Recalls (openFDA)

Instrumentation Laboratory: GEM Premier 5000 PAK Part Numbers: 00055360010 00055360011 00055407504 00055415010 00055430010 00055430011 00055445010 00055445011 00055360010.A

Recall number
Z-2900-2026
Recalling firm
Instrumentation Laboratory
Product
GEM Premier 5000 PAK Part Numbers: 00055360010 00055360011 00055407504 00055415010 00055430010 00055430011 00055445010 00055445011 00055360010.A
Status
Open, Classified
Initiated
2026-05-21
Posted
2026-08-06
Terminated
not on file
Reason
The active ingredient concentration in the lysing bag solution of specific GEM Premier 5000 lots did not meet established manufacturing specifications. The malfunction could result in delayed recognition and management of Co-Oximetry and total bilirubin abnormalities and their sequelae.
Root cause
Nonconforming Material/Component
Action
On 5/21/26, the firm began notifying consignees in the US via letter. Consignees were informed that Werfen will be proactively shipping replacement PAKs to each facility. Customers were instructed to check inventory for affected product. They should discontinue use of the affected lots and destroy any inventory on hand. Destruction should be documented via the response form provided to the customer and returned to the recalling firm. Customers should forward the notification to all affected locations at each facility. On May 28, 2026, the field action was expanded to include affected customers in China. On June 23, 2026, the field action was expanded to include additional affected lots in Italy and Germany.
Quantity
109 units
Distribution
Worldwide - US Nationwide distribution in the states of AZ, CA, FL, GA, IL, IN, KY, MN, MO, MT, OH, PA, VA, WI, WV and the countries of China, Germany, Italy.
Lot, serial or model codes
Part No. 00055360010 UDI-DI 08426950807599 Lots 260505K, 260504AG, 260505F; Part No. 00055360011 UDI-DI 08426950807605 Lots 260505F and 260505J; Part No. 00055407504 UDI-DI 08426950807537 Lot 260505R; Part No. 00055415010 UDI-DI 08426950807698 Lots 260505H and 260505K; Part No. 00055430010 UDI-DI 08426950807759 Lots 260505H, 260505K and 260505F; Part No. 00055430011 UDI-DI 08426950807766 Lot 260505K; Product No. 00055445010 UDI-DI 08426950807810 Lots 260504AA and 260505K; Product No. 00055445011 UDI-DI 08426950807827 Lots 260505F and 260505J; Product No. 00055360010.A UDI-DI 08426950807599 Lot 260504AF;
510(k) numbers
["K203790"]
PMA numbers
not on file
Product code
CHL
Firm FEI
1217183
Firm city
Bedford
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register