Skip to the content

FDA Medical Device Recalls (openFDA)

Axiobionics, LLC: AxioBionics Wearable Therapy Upper Extremity BioSleeve (Right and Left) that includes 2-16 BioGel Velcro Electrodes in sizes: 2"x2", 2"x2.5", 2"x3", 2"x3.5", 2"x4", 2"x5", 2"x6", 2"x8", 1.5"x3", 1.5"x5", 1.5"x6", and 1"x1.5". Cutaneous electrode garments for neuromuscular electrical stimulation.

Recall number
Z-2898-2011
Recalling firm
Axiobionics, LLC
Product
AxioBionics Wearable Therapy Upper Extremity BioSleeve (Right and Left) that includes 2-16 BioGel Velcro Electrodes in sizes: 2"x2", 2"x2.5", 2"x3", 2"x3.5", 2"x4", 2"x5", 2"x6", 2"x8", 1.5"x3", 1.5"x5", 1.5"x6", and 1"x1.5". Cutaneous electrode garments for neuromuscular electrical stimulation.
Status
Terminated
Initiated
2011-05-27
Posted
2011-08-01
Terminated
2013-08-29
Reason
Premature Device Failure : Failure of the electrode sites to conduct and provide muscle stimulation due to silver fabric component failure.
Root cause
Nonconforming Material/Component
Action
Axiobionics, LLC is notifying all affected customers by phone. Three attempts will be made to contact the customer by phone. Contact by telephone will be made to prescribing physician to determine if Axiobionics has latest patient contact information. If information is different, new contact information will be recorded and 3 additional attempts will be made with the new information. If the patient cannot be reached by telephone, a certified letter will be sent to the patient's address. If contacting by mail, the letter template will be provided to the FDA recall office . All patients should have been contacted by telephone, certified mail, or in person by June 10, 2011. All patients affected by the field correction will be asked the following questions to determine if their electrodes are functional and resistances are not exceeding the output capabilities of the stimulator they are using. 1. Are you using your garment? 2. Is your garment delivering the stimulation required for your therapy? 3. If the patient is using an NP4 stimulator: What are the resistance readings on each channel? 4. If the patient answers No to question 2 or states that the resistance is reading greater than 200 ohms on at least one channel in question 3, the garments will be returned to Axiobionics for repair. Garments that are not functioning correctly will be repaired first. Functioning garments will be tested by immersion into a salt bath under the given protocol to determine if the silver fabric is properly coated. Questions regarding this correction should be directed to the Axiobionics representative at 800-552-3539.
Quantity
35
Distribution
Nationwide Distribution--USA (nationwide) including the states of AZ, GA, IL, IN, MI, MO, NY, TX, and WY.
Lot, serial or model codes
Catalog numbers for BioSleeve include 10200 and 10201 and Serial numbers 1466-1469, 1472, 1380, 1425, 1371-1374, 1389, 1391, 1392, 1453, 1454, 1460, 1393, 1402, 1403, 1430, 1482, 1396, 1502, 1503, 1431, 1432, 1449, 1435, 1387, 1504, 1409, 1410, 1539 and 1540. Catalog numbers for the BioGel Velcro Electrodes include118, 119, 120, 121, 122, 123, 124, 125, 126, 127, 128, 129, and 130.
510(k) numbers
["K944543"]
PMA numbers
not on file
Product code
GXY
Firm FEI
3008525454
Firm city
Ann Arbor
Firm state
MI
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register