Bio-Rad Laboratories Inc: D-10 Dual Program Reorder Pack; contains 400 HbA1c or 200 HbA2/F/A1c tests. Kits include Elution Buffer 1 and 2; Calibrator 1 & 2. Bio-Rad Laboratories Inc. Hercules, CA. Product Usage: Short Proghram: D-10 Dual Program is intended for the percent determination of hemoglobin A1c in human whole blood using ion-exchange high-performance liquid chromatography (HPLC). Extended Program: The Bio-Rad D-10 Dual Program is intended for the percent determination of hemoglobins A2, F, and A1c, and for the detection of abnormal hemoglobins in human whole blood using ion-exchange high-performance liquid chromatography (HPLC). The Bio-Rad D-10 Dual Program is intended for Professional Use Only: For In-Vitro Diagnostic Use.
- Recall number
- Z-2889-2011
- Recalling firm
- Bio-Rad Laboratories Inc
- Product
- D-10 Dual Program Reorder Pack; contains 400 HbA1c or 200 HbA2/F/A1c tests. Kits include Elution Buffer 1 and 2; Calibrator 1 & 2. Bio-Rad Laboratories Inc. Hercules, CA. Product Usage: Short Proghram: D-10 Dual Program is intended for the percent determination of hemoglobin A1c in human whole blood using ion-exchange high-performance liquid chromatography (HPLC). Extended Program: The Bio-Rad D-10 Dual Program is intended for the percent determination of hemoglobins A2, F, and A1c, and for the detection of abnormal hemoglobins in human whole blood using ion-exchange high-performance liquid chromatography (HPLC). The Bio-Rad D-10 Dual Program is intended for Professional Use Only: For In-Vitro Diagnostic Use.
- Status
- Terminated
- Initiated
- 2011-07-11
- Posted
- 2011-07-27
- Terminated
- 2012-03-12
- Reason
- Reports of calibration failure and late retention time using Elution Buffer AA10434 and AA10435.
- Root cause
- Other
- Action
- Bio-Rad Laboratories Inc sent "URGENT: MEDICAL DEVICE CORRECTION" letters to all affected D-10 Dual customers and Bio-Rad subsidiaries on July 11, 2011 via FedEx or fax. The letters identified the product, the problem, and the action needed to be taken. The letters ask that inventory be examined and affected reorder pack lots be quarantined and destroyed according to local, state and federal regulations. All D-10 Dual customers were instructed to review their chromatograms generated from these lots. If the calibration has passed with good chromatography, the controls are within specification, and the F, A1c, and A2 peaks elute in their specified retention time windows for the patient samples, the customer may continue to use and report results with the affected buffer lots. The letters include a chart showing Good Patient Chromatogram with Correct Retention times and Failed Calibrator Chromatogram with late retention times. If calibration fails and/or late retention times cause one or more peaks to be misidentified, affected reorder packs should be destroyed. The correction letters and response form should be translated, if necessary, and distributed to all affected customers. Tracking documentation should be compiled for all customers who receive the Medical Device Correction letter. The Medical Device Correction Response Forms should be completed and returned by fax to Bio-Rad CSD Regulatory Affairs at 510-741-3954 or a PDF copy to Jackie_Buckley@bio-rad.com
- Quantity
- 418
- Distribution
- Worldwide -- USA (nationwide) including the states of: AR, CA, CT, FL, HI, IN, MD, MN, MO, MS, NJ, NY, OH, PA, RI, TX. and the countries of: Brazil, Canada, China, England, France, India, Italy, Mexico, New Zealand, and S. Korea.
- Lot, serial or model codes
- Model number 220-0201; Kit Lot number/Expiration date: 20010663/Sept 2012; 20010665-A/Sept 2012; 20010665-B/Sept 2012; 20010665-C/Sept 2012. Elution Buffer Lot numbers: AA10434, AA10435; Exp 9/30/2012
- 510(k) numbers
- ["K041444"]
- PMA numbers
- not on file
- Product code
- LCP
- Firm FEI
- 2915274
- Firm city
- Hercules
- Firm state
- CA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28