Civco Medical Instruments Co. Inc.: CIVCO lnfiniti Needle Guide Bracket. used for performing needle-guided (or catheter) procedures with the use of diagnostic ultrasound transducers. Product Usage: The bracket and disposable guide provide physicians with a tool for performing needle-guided (or catheter) procedures with the use of diagnostic ultrasound transducers.

Recall number
Z-2888-2018
Recalling firm
Civco Medical Instruments Co. Inc.
Product
CIVCO lnfiniti Needle Guide Bracket. used for performing needle-guided (or catheter) procedures with the use of diagnostic ultrasound transducers. Product Usage: The bracket and disposable guide provide physicians with a tool for performing needle-guided (or catheter) procedures with the use of diagnostic ultrasound transducers.
Status
Terminated
Initiated
2018-02-23
Posted
not on file
Terminated
2020-10-06
Reason
A cover failed leak testing during validation testing of new needle guides from a replacement tool.
Root cause
Device Design
Action
Telephone Script to be read: CIVCO is calling to notify our customers regarding a recall action being initiated. CIVCO has recently discovered a dimensional issue in our Infiniti bracket which may allow the needle guide to a puncture a cover. You are being notified, as your facility has been identified as having received product from one of the affected manufacturing lots. Advise on Action to be taken by Customer or Distributor: To return Infiniti needle guide, contact your Account Representative at 800-445-6741 or 319-248-6757. Alternately you may also use email at order@civco.com to schedule the return of your device. Additional information to obtain: Verification of Email contact to send notification letter and response letter.
Quantity
177 boxes
Distribution
Worldwide Distribution - US Nationwide in the states of GA, KY, CA, NY , NJ, TX, IL, MA, RI, AZ, and IA and countries of Ukraine, Hong Kong, Italy, Netherlands, Switzerland, United Kingdom, Denmark, Germany, Belgium, Spain, Japan, China, and Canada
Lot, serial or model codes
Item: 639-026 - Biopsy Starter Kit for use with Esaote LA332 and SL332 transducers Lot #'s M206950, M664690, M897110, M119380, M605540, M706720, M570610, M543670, M851530, M983170, M358570. M358570, M427660. M526300, M526300, M734570, M912740, and M664690. Item 698-009: Biopsy Starter Kit for use with Zonare C9-3 transducers M427370, M103030, A010473, M103030, M771190, M427390, M293590, M427370, M105150, M089260, M515370, M293580, M888920, M427370, M220960, M558420, M560740, M909270, M096330, M388690, M558220, M695400, M085550, M917410, M419650, M879530, M964370, M093020, M723180, and A010459 Item: 698-016 Biposy Bracket for use with Zonare C9-3 transducers M567620 and M052860 Item 620-116 Biopsy Starter Kit for use with BK Medical 18L5 and 8870 transducers M080890, M559550, M425050, M056610, M092800, M041630, M186540, M122250, M335740, M043310, M499390, M483250, M109060, M275630, M912150, A023904, M457560, M051670, M091360, M895560, A015235, M202260, M869430, M056610, M027090, M308820, M831280, M547370, M253050, M083160, and M865470
510(k) numbers
["K093713"]
PMA numbers
not on file
Product code
ITX
Firm FEI
1937223
Firm city
Kalona
Firm state
IA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register