Spiegelberg Gmbh & Co. KG: Spiegelberg Brain Pressure Monitoring System Probe 3PN, REF: SNDI3.1.531FV534P and Probe 3PS, REF: SNDI3.1.63/FV535P Product Usage: lntracranial pressure measuring probes are used to measure intracranial pressure (intracranial pressure, ICP), so the pressure in the head.

Recall number
Z-2888-2016
Recalling firm
Spiegelberg Gmbh & Co. KG
Product
Spiegelberg Brain Pressure Monitoring System Probe 3PN, REF: SNDI3.1.531FV534P and Probe 3PS, REF: SNDI3.1.63/FV535P Product Usage: lntracranial pressure measuring probes are used to measure intracranial pressure (intracranial pressure, ICP), so the pressure in the head.
Status
Terminated
Initiated
2009-12-29
Posted
not on file
Terminated
2016-10-05
Reason
In the vast majority of products specified, a shrinkage of the air chamber during sterilization has occurred. Because of these false high readings were displayed in the lower measuring range from 0 to 20 mmHg. The observed measurement error was particularly at the lowest end of the range (to 0 mmHg) up to 8 mmHg.
Root cause
Packaging
Action
Spiegelberg Gmbh & Co. KG sent an Urgent Medical Device Recall letter dated December 29, 2009, to all affected customers. Customers were instructed to quarantine and return stored products. Furthermore, they were asked to inform all their customers which have received the affected products and recall the products accordingly. Furthermore, they were asked to tell us about the actions taken by them.
Quantity
22 units
Distribution
Nationwide Distribution to PA only
Lot, serial or model codes
Probe 3PN SND1 3.1 .53/FV534P: SNs: 12835-1 3393;  Probe 3PS SND13.1 .63/FV535P: SNs: 6937-7122
510(k) numbers
["K003759"]
PMA numbers
not on file
Product code
GWM
Firm FEI
3002108226
Firm city
Hamburg
Firm state
not on file
Firm country
Germany

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register