Arrow International Inc: Pediatric Two-Lumen Central Venous Catheterization Kit with Blue FlexTip ARROWg+ard Blue Catheter, Cat. No. AK-22502 Product Usage: The multiple-lumen catheter permits venous access to the central circulation of pediatric patients. The ARROWg+ard antimicrobial surface is intended to help provide protection against catheterrelated infections. The catheter is not intended to be used as a treatment for existing infections, nor is it indicated for long-term use. The use of the 4, 5, and 5.5 Fr. antimicrobial surface catheters in infants weighing less than 2.0, 2.5, and 3.0 kilograms respectively, have not been evaluated. Therefore, the benefits of these catheters in infants weighing less than the indicated weights should be weighed against any possible risks.

Recall number
Z-2886-2018
Recalling firm
Arrow International Inc
Product
Pediatric Two-Lumen Central Venous Catheterization Kit with Blue FlexTip ARROWg+ard Blue Catheter, Cat. No. AK-22502 Product Usage: The multiple-lumen catheter permits venous access to the central circulation of pediatric patients. The ARROWg+ard antimicrobial surface is intended to help provide protection against catheterrelated infections. The catheter is not intended to be used as a treatment for existing infections, nor is it indicated for long-term use. The use of the 4, 5, and 5.5 Fr. antimicrobial surface catheters in infants weighing less than 2.0, 2.5, and 3.0 kilograms respectively, have not been evaluated. Therefore, the benefits of these catheters in infants weighing less than the indicated weights should be weighed against any possible risks.
Status
Terminated
Initiated
2018-07-03
Posted
not on file
Terminated
2020-04-29
Reason
One lot of product is from product code AK-22502 so it contains a 2-lumen, 5 Fr x 5 cm catheter. However, this product is incorrectly labeled as AK-25502, so the lidstock says it contains a 2-lumen, 5 Fr x 8 cm catheter. No patient injuries have been reported related to this issue.
Root cause
Employee error
Action
On July 3, 2018, the firm sent out Urgent Medical Device Recall letters to customers. The letter communicated the labeling error. Affected consignees are instructed to take the following actions: 1. If you have affected stock, immediately discontinue use and quarantine any products with the product code and lot number listed above, so that the affected products can be returned to Arrow. 2. To return product, complete the enclosed Recall Acknowledgement Form and fax it to 1-855-419-8507, Attn: Customer Service or email to recalls@teleflex.com. This will allow us to document the amount of product you have on hand for return. A customer service representative will contact you with a Return Goods Authorization (RGA) Number and will provide instructions for the return of products to Arrow International. 3. If you have no affected stock, please complete the enclosed Recall Acknowledgment Form and fax it to 1-855-419-8507, Attn: Customer Service or email to recalls@teleflex.com. This will allow us to document your receipt of this letter.
Quantity
166
Distribution
Distributors in MO and UT. Medical facilities in AL, CA, IL, MA, NH, NM, NY, TX, VA, and WI.
Lot, serial or model codes
13F17K0202
510(k) numbers
["K900263"]
PMA numbers
not on file
Product code
FOZ
Firm FEI
3015859709
Firm city
Reading
Firm state
PA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register