Skip to the content

FDA Medical Device Recalls (openFDA)

The Magstim Company Limited: Neurosign V4 Intraoperative Nerve monitor family of devices.

Recall number
Z-2884-2020
Recalling firm
The Magstim Company Limited
Product
Neurosign V4 Intraoperative Nerve monitor family of devices.
Status
Terminated
Initiated
2020-06-23
Posted
not on file
Terminated
2024-08-16
Reason
Reports that manipulation of the cable which connects the pre-amplifier to the nerve monitor, and the cable which connects the pre-amplifier to the stimulator pod, may cause damage to the internal construction of the cables.
Root cause
Device Design
Action
On June 23 2020, the firm sent a urgent field safety notice to its costumers. What you need to do: - Identify all affected products at your facility - Raise awareness to all users of the device(s) within your organisation - If you have forwarded any affected product to other organisations, please forward them a copy of this notice, and advise us accordingly on the acknowledgement form - Complete and return the attached acknowledgement form, so we can contact you to arrange replacement of the cable. - Users should continue to routinely inspect all devices for damage or signs of wear before use, as described in the operating manual. - If damage is found or your device is exhibiting the failure modes described, contact the service department to arrange a repair or replacement. - If damage is not found and the device is behaving as expected, the device can continue to be used whilst a replacement is arranged. Care should be taken to properly store the Pre- Amplifier and Stimulator Pod using the cable storage aids provided. What happens next? The affected cables have been re-designed to improve the longevity and cable performance. Using the details provided on the acknowledgement form, we will arrange for a replacement of affected cables at your facility to the updated design. It is important to return the acknowledgement form as soon as possible so that we can arrange for your cables to be replaced. In the meantime, you can continue to use your device(s).
Quantity
49 devices
Distribution
Domestic: MN International: UK, Republic of Ireland, Spain, Netherlands, Canada. (Technomed Europe distributed to: Brazil, Spain, Belgium, Chile, Greece, Poland, New Zealand, Pakistan, Portugal and Thailand.
Lot, serial or model codes
Serial numbers # 001 to 049.  US product serial number is 039.
510(k) numbers
["K181559"]
PMA numbers
not on file
Product code
PDQ
Firm FEI
3002808356
Firm city
Whitland
Firm state
not on file
Firm country
United Kingdom

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register