FDA Medical Device Recalls (openFDA)

Beckman Coulter Inc.: DHEA RIA Reagent The DSL8900 DHEA radioimmunoassay is intended for quantitative determination of dehydroepiandrosterone in serum or plasma. This assay is intended for in vitro diagnostic use.

Recall number
Z-2882-2011
Recalling firm
Beckman Coulter Inc.
Product
DHEA RIA Reagent The DSL8900 DHEA radioimmunoassay is intended for quantitative determination of dehydroepiandrosterone in serum or plasma. This assay is intended for in vitro diagnostic use.
Status
Terminated
Initiated
2010-09-09
Posted
2011-07-26
Terminated
2012-05-04
Reason
The recall was initiated because Beckman Coulter has confirmed customer reports that DHEA measurements using DHEA RIA (REF DSL-8900) kit lots identified may overstate actual DHEA levels in patient samples.
Root cause
Process control
Action
Beckman Coulter sent an "URGENT: PRODUCT CORRECTION ACTION" letter dated November 15, 2010 to all affected customers. The letter describes the product, problem, and actions to be taken by the customers. The letter instructs customers to discontinue using the affected products and discard them according to local regulations. Customers were asked to review patient DHEA results ran on the affected lots and interpret them in light of additional tests (such as LH, FSH, testosterone, DHEA-S, cortisol, 17-hydroxyprogesterone, and estradiol) and the patient's clinical status. A response form was enclosed for customers to complete and return. Contact the Beckman Coulter Technical Support Center at 1-800-854-3633 in the United States, or their local Beckman Coulter Representative for questions regarding this notification.
Quantity
3,542 units (1181 in US)
Distribution
Worldwide Distribution--USA (nationwide) and the countries of Argentina, Australia, Belgium, Brazil, Canada, Chile, China, Czech Republic, France, Germany, Hungary, India, Ireland, Italy, Japan, Mexico, Netherlands, Poland, Spain, Switzerland, Taiwan, Turkey, and United Kingdom.
Lot, serial or model codes
Lot numbers: 991112, 991346, 991584, 991857, 992115, 992359, 992458, 992854 were manufactured at: Beckman Coulter, Diagnostics Systems Laboratories, a Beckman Coulter Company, 445 West Medical Center Blvd, Webster, Texas 77598.  Lot numbers: 100115A, 100205A, 100205RA, 100226A, 100226RA, 100319A, 100319RA, 100409A, 100409RA, 100430A, 100430RA, 100521A, 100611A, 100611RA, 100702A, 100723A, 100723B, 100723RA, 100813A, 100813RB, 100813C were manufactured at: Immunotech, Prague, a Beckman Coulter Company, Radiova 1, Praha, 10227, Czech Republic.
510(k) numbers
["K921534"]
PMA numbers
not on file
Product code
JKC
Firm FEI
2050012
Firm city
Brea
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register