FDA Medical Device Recalls (openFDA)
In2bones USA, LLC: CoLink(TM) Lapidus Plate XP, +2 mm, Left, REF P40 ST265, QTY 1, In2Bones Product Usage: The In2Bones USA LLC, CoLink Plating System is indicated for stabilization and fixation of fractures, revision procedures, joint fusion, osteotomies and reconstruction of the small bones in the hand, wrist, foot and ankle in both pediatric and adult patients.
- Recall number
- Z-2880-2018
- Recalling firm
- In2bones USA, LLC
- Product
- CoLink(TM) Lapidus Plate XP, +2 mm, Left, REF P40 ST265, QTY 1, In2Bones Product Usage: The In2Bones USA LLC, CoLink Plating System is indicated for stabilization and fixation of fractures, revision procedures, joint fusion, osteotomies and reconstruction of the small bones in the hand, wrist, foot and ankle in both pediatric and adult patients.
- Status
- Terminated
- Initiated
- 2018-07-16
- Posted
- not on file
- Terminated
- 2019-02-22
- Reason
- CoLink(TM) Lapidus Plate +2mm Right and Left, were laser marked incorrectly. In these lots, right plates were laser marked as left plates and the left plates were laser marked as right plates.
- Root cause
- Process control
- Action
- In2Bones sent a Medical Device Recall letter dated July 18, 2018 to affected customers (agents and hospitals). The letter identified the affected product, problem and actions to be taken. The Agent notice requested the consignee to inform and distribute the notice to all relevant person within the organization. The firm is seeking the return of the products The notice to the hospitals and healthcare professionals explained the problem and recommended that patients be monitored for any adverse events within the normal standard of practice. For questions contact Quality and Regulatory Affairs Team at 901-260-7931..
- Quantity
- 48 units
- Distribution
- Worldwide Distribution - US Nationwide and the countries of France
- Lot, serial or model codes
- Lot Number 1704075
- 510(k) numbers
- ["K163293"]
- PMA numbers
- not on file
- Product code
- HWC
- Firm FEI
- 3011580264
- Firm city
- Memphis
- Firm state
- TN
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28