Skip to the content

FDA Medical Device Recalls (openFDA)

CareFusion 303, Inc.: Alaris Syringe Module Model 8110 with software version 9.15 The Alaris Syringe Pump module is part of the Alaris System. The syringe pump delivers fluids in a manner similar to current syringe pumps on the market. Up to four Alaris Syringe pump modules can be connected to the Alaris PC unit which is the central programming, monitoring and power supply component for the Alaris System. The syringe pump uses standard, single-use administration sets and syringes with luer-lock connectors.

Recall number
Z-2879-2016
Recalling firm
CareFusion 303, Inc.
Product
Alaris Syringe Module Model 8110 with software version 9.15 The Alaris Syringe Pump module is part of the Alaris System. The syringe pump delivers fluids in a manner similar to current syringe pumps on the market. Up to four Alaris Syringe pump modules can be connected to the Alaris PC unit which is the central programming, monitoring and power supply component for the Alaris System. The syringe pump uses standard, single-use administration sets and syringes with luer-lock connectors.
Status
Terminated
Initiated
2016-08-08
Posted
not on file
Terminated
2018-10-26
Reason
A software anomaly with the Alaris Syringe module software version 9.15 may cause an infusion to unexpectedly stop when the Syringe module is transitioning from one rate to another.
Root cause
Software design
Action
The firm, BD, sent an "Urgent: Medical Device Recall Notification" letter dated 8/8/16 to customers to inform them that BD has identified a software anomaly associated with the Alaris Syringe module software version 9.15. The letter describes the product, problem and actions to be taken. The customers were instructed to follow the instructions and promptly complete and return the Customer Response Card to expedite the corrective action process. Customers with recall related questions are instructed to contact the BD Support Center at (888) 562-6018. Customers with adverse event reports are instructed to contact customer advocacy at (888) 812-3266. Customers with technical questions regarding the Alaris System are instructed to contact Technical Support at (888) 812-3229.
Quantity
12,000 units
Distribution
Worldwide Distribution: US(nationwide) and countries of: Canada, Australia and United Arab Emirates.
Lot, serial or model codes
Software version 9.15
510(k) numbers
["K023264"]
PMA numbers
not on file
Product code
FRN
Firm FEI
2016493
Firm city
San Diego
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register