Skip to the content

FDA Medical Device Recalls (openFDA)

Merge Healthcare, Inc.: Merge PACS, medical imaging diagnostic workstation

Recall number
Z-2874-2026
Recalling firm
Merge Healthcare, Inc.
Product
Merge PACS, medical imaging diagnostic workstation
Status
Open, Classified
Initiated
2026-06-30
Posted
2026-08-04
Terminated
not on file
Reason
Merge Healthcare has identified a potential issue that may cause thumbnails from a previously opened patient study to be displayed in the Thumbnail Bar while a different patient study is open.
Root cause
Software Design Change
Action
Merge Healthcare issued an Urgent Field Safety Notice (Revision 2) to its consignees on 6/30/2026 via email. The notice explained the issue and when it may occur, risk to patient, and requested an immediate workaround, a configuration change, be performed to disable automatic refresh of the Thumbnail Bar until a software update is available. The notice also directions the consignee to forward the notice to any individuals within their organization that may need to be aware of this notification. Customers should alert any other affiliated parties that may be affected by this Urgent Field Safety Notice. The notice was followed by telephone calls beginning on 7/6/2026 to ensure the notice was received. For more information, please contact Merge Healthcare Support at 1-877-741-5369 (available 24 hours a day, 7 days a week) or visit our website: https://www.merative.com/support/merge.
Quantity
53 units
Distribution
Worldwide distribution - US Nationwide and the countries of Canada, and UK.
Lot, serial or model codes
UDI (01)00842000101215(10)9.4.0.P01(11)250916, (01)00842000101154(10)9.4.0.P02(11)260302 with Affected Versions: 9.4, 9.4 P01, and 9.4 P02.
510(k) numbers
["K192455"]
PMA numbers
not on file
Product code
LLZ
Firm FEI
1000116130
Firm city
Hartland
Firm state
WI
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.