GE Healthcare Japan Corporation: GE Healthcare Optima CT660. The systems are intended for head, whole body, cardiac and vascular X-ray Computed Tomography applications. Sold only outside the U.S. under the following product names: CT OPTIMA 660, CT OPTIMA 660 LONG BJG SKD RUS, CT OPTIMA 660 LONG BJG SKD RUSSIA, CT OPTIMA 660 MID BJG SKD RUSSIA, CT OPTIMA 660 SE, CT OPTIMA 660 SE ASIR, CT660 M40 LONG, CT660 M40 MID, CT660 M40 MID WAUK SKD, CT660 M40 MID WAUK SKD BRAZ, CT660 M40 MID WAUK SKD BRAZ, OPT 660 M40 MID HINO SKD RUS, OPTIMA 660, OPTIMA 660 1.5, OPTIMA 660 1.5 BRAZIL, OPTIMA 660 1.5 BRAZIL SKD W, OPTIMA 660 1.5 BRAZIL SKD WAUK, OPTIMA 660 LONG BJG, OPTIMA 660 M40, OPTIMA 660 M40 3.5 LONG BJG, OPTIMA 660 M40 LONG BJG, OPTIMA 660 M40 LONG BJG SKD RUSS, OPTIMA 660 M40 MID BJG, OPTIMA 660 M40 MID BJG SKD RUSS, OPTIMA 660 MID BJG, OPTIMA 660 MID BJG SKD, OPTIMA CT 660, OPTIMA CT660, OPTIMA CT660 ASIR 128 M RU, OPTIMA CT660 FREEDOM LG HN2.5, OPTIMA CT660 FREEDOM MD BJG2.5, OPTIMA CT660 MID P1.5 HINO, OPTIMA CT660 SE ASIR, OPTIMA CT660 SE GT 1700 HN2.5, OPTIMA CT660 SE GT 2000 BJG2.5, OPTIMA CT660 SKD BRAZIL, OPTIMA CT660S L RU
- Recall number
- Z-2873-2018
- Recalling firm
- GE Healthcare Japan Corporation
- Product
- GE Healthcare Optima CT660. The systems are intended for head, whole body, cardiac and vascular X-ray Computed Tomography applications. Sold only outside the U.S. under the following product names: CT OPTIMA 660, CT OPTIMA 660 LONG BJG SKD RUS, CT OPTIMA 660 LONG BJG SKD RUSSIA, CT OPTIMA 660 MID BJG SKD RUSSIA, CT OPTIMA 660 SE, CT OPTIMA 660 SE ASIR, CT660 M40 LONG, CT660 M40 MID, CT660 M40 MID WAUK SKD, CT660 M40 MID WAUK SKD BRAZ, CT660 M40 MID WAUK SKD BRAZ, OPT 660 M40 MID HINO SKD RUS, OPTIMA 660, OPTIMA 660 1.5, OPTIMA 660 1.5 BRAZIL, OPTIMA 660 1.5 BRAZIL SKD W, OPTIMA 660 1.5 BRAZIL SKD WAUK, OPTIMA 660 LONG BJG, OPTIMA 660 M40, OPTIMA 660 M40 3.5 LONG BJG, OPTIMA 660 M40 LONG BJG, OPTIMA 660 M40 LONG BJG SKD RUSS, OPTIMA 660 M40 MID BJG, OPTIMA 660 M40 MID BJG SKD RUSS, OPTIMA 660 MID BJG, OPTIMA 660 MID BJG SKD, OPTIMA CT 660, OPTIMA CT660, OPTIMA CT660 ASIR 128 M RU, OPTIMA CT660 FREEDOM LG HN2.5, OPTIMA CT660 FREEDOM MD BJG2.5, OPTIMA CT660 MID P1.5 HINO, OPTIMA CT660 SE ASIR, OPTIMA CT660 SE GT 1700 HN2.5, OPTIMA CT660 SE GT 2000 BJG2.5, OPTIMA CT660 SKD BRAZIL, OPTIMA CT660S L RU
- Status
- Terminated
- Initiated
- 2018-04-27
- Posted
- not on file
- Terminated
- 2020-08-21
- Reason
- GE Healthcare has identified that some CT systems may have a damaged cable that can expose 120VAC to Service Engineers working inside the gantry.
- Root cause
- Employee error
- Action
- The firm, GE Healthcare, sent an " URGENT MEDICAL DEVICE CORRECTION" letter dated 4/27/2018 to its Customers on about 04/27/2018. The letter describes the product, problem and actions to be taken. The customers were Instructed to: download and provide the updated Service Manual to the organization that services the equipment on site and informed customers that GE Healthcare would be correcting all affected devices at no cost. Questions or concerns can be directed to GE Healthcare Service at 1-800-437-1171 or the local Sales Representative. GE Healthcare expanded the recall to include an additional product and model number. Customers were notified on about 08/09/2018 via URGENT MEDICAL DEVICE CORRECTION letter. Instructions were the same as the original notification.
- Quantity
- 92 OUS only
- Distribution
- Worldwide Distribution: US (Nationwide) distribution to states of: AK, AL, CA, CO, CT, DE, FL, GA, GU, HI, IA, ID, IL, IN, KS, KY, LA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, RI, SC, SD, TN, TX, UT, WI, WV, WY; and countries of:: ALGERIA, ARGENTINA, ARMENIA, AUSTRALIA, AUSTRIA, AZERBAIJAN, BAHRAIN, BELARUS, BELGIUM, BOLIVIA, BOSNIA AND HERZEGOVINA, BRAZIL, BULGARIA, CAMEROON, Canada, CHILE, CHINA, COLOMBIA, Cote D'Ivoire, CROATIA, CYPRUS, CZECH REPUBLIC, DENMARK, ECUADOR, EGYPT, ETHIOPIA, FINLAND, FRANCE, French Guiana, GERMANY, GHANA, GREECE, GUADELOUPE, GUATEMALA, HONDURAS, Hong Kong, HUNGARY, INDIA, INDONESIA, IRAN, IRAQ, IRELAND, ISRAEL, ITALY, JAPAN, JORDAN, KAZAKHSTAN, Korea (Republic of), KUWAIT, LATVIA, LEBANON, LIBYA, LITHUANIA, MALAYSIA, Mali, MALTA, MARTINIQUE, MEXICO, MOROCCO, NETHERLANDS, NEW , ZEALAND, NIGERIA, NORWAY, OMAN, PAKISTAN, Palestine, PANAMA, PERU, PHILIPPINES, POLAND, PORTUGAL, PUERTO RICO, REUNION, ROMANIA, Russia, RWANDA, SAUDI ARABIA, SINGAPORE, SLOVAKIA, SLOVENIA, SOUTH AFRICA, SPAIN, SWEDEN, SWITZERLAND, TAIWAN, TANZANIA, THAILAND, TUNISIA, TURKEY, UKRAINE, UNITED ARAB EMIRATES, UNITED KINGDOM, URUGUAY, UZBEKISTAN, VENEZUELA, VIETNAM, and YEMEN.
- Lot, serial or model codes
- Model Number, Serial Number (System ID): 6375965-146, CJRSX1600007CN (CJRSX1600007CN); 6375965-146, CJRSX1600001CN (CJRSX1600001CN); 6375965-100, 00000433543CN6 (CT433543CN6); 6375965-100, 00000438341CN0 (CT438341CN0); 6375965-146, CJRSX1600003CN (CJRSX1600003CN); 6375965-146, CJRSX1600004CN (CJRSX1600004CN); 6375965-146, CJRSX1700005CN (CJRSX1700005CN); 6375965-146, CJRSX1700009CN (CJRSX1700009CN); 6375965-100, 00000432220CN2 (CT432220CN2); 6375965-100, 00000438956CN5 (CT438956CN5); 6375965-146, CJRSX1700012CN (CJRSX1700012CN); 6375965-100, 00000438278CN4 (CT438278CN4); 6375965-100, 00000439130CN6 (CT439130CN6); 6375965-146, CJRSX1700003CN (CJRSX1700003CN); 6375965-100, 00000433540CN2 (CT433540CN2); 6375965-100, 00000440700CN3 (CT440700CN3); 6375965-100, 00000441516CN2 (CT441516CN2); 6375965-100, 00000431824CN2 (CT431824CN2); 6375965-146, CJRSX1700004CN (CJRSX1700004CN); 6375965-100, 00000426748CN0 (CT426748CN0); 6375965-100, 00000435884CN2 (CT435884CN2); 6375965-146, CJRSX1600002CN (CJRSX1600002CN); 6375965-100, 00000440598CN1 (CT440598CN1); 6375965-146, 00000443895CN8 (CT443895CN8); 6375965-100, 00000438072CN1 (CT438072CN1); 6375965-100, 00000438426CN9 (CT438426CN9); 6375965-100, 00000441729CN1 (CT441729CN1); 6375965-100, 00000439438CN3 (CT439438CN3); 6375965-100, 00000437682CN8 (CT437682CN8); 6375965-100, 00000441405CN8 (CT441405CN8); 6375965-100, 00000426912CN2 (CT426912CN2); 6375965-100, 00000441984CN2 (CT441984CN2); 6375965-100, 00000432339CN0 (CT432339CN0); 6375965-100, 00000433108CN8 (CT433108CN8); 6375965-146, CJRSX1500002CN (CJRSX1500002CN); 6375965-100, 00000431520CN6 (CT431520CN6); 6375965-100, 00000440381CN2 (CT440381CN2); 6375965-100, 00000426771CN2 (CT426771CN2); 6375965-100, 00000437987CN1 (CT437987CN1); 6375965-146, CJRSX1600006CN (CJRSX1600006CN); 6375965-100, 00000435820CN6 (CT435820CN6); 6375965-100, 00000437399CN9 (CT437399CN9); 6375965-100, 00000439050CN6 (CT439050CN6); 6375965-100, 00000439437CN5 (CT439437CN5); 6375965-100, 00000439974CN7 (CT439974CN7); 6375965-100, 00000439998CN6 (CT439998CN6); 6375965-146, CJRSX1500001CN (CJRSX1500001CN); 6375965-146, CJRSX1700011CN (CJRSX1700011CN); 6375965-100, 00000432173CN3 (CT432173CN3); 6375965-203, 00000306435HM9 (RU2116CT02); 6375965-6 , RSGNA1700007YC (RU4417CT01); 6375965-203, 00000314586HM9 (RU2117CT02); 6375965-203, 00000305388HM1 (RU1084CT02); 6375965-6 , RSGNA1700004YC (RU0191CT01); 6375965-203, 00000342945HM3 (RU7346CT01); 6375965-203, 00000342946HM1 (RU7635CT01); 6375965-203, 00000366681HM5 (RU1254CT03); 6375965-203, 00000385668HM9 (RU2609CT01); 6375965-203, 00000387806HM3 (RU9059CT01); 6375965-6 , RSGNA1700003YC (RU1581CT01); 6375965-6 , RSGNA1700008YC (RU1928CT01); 6375965-6 , 00000069003YC2 (RU8267CT01); 6375965-6 , 00000077301YC0 (RU1107CT02); 6375965-6 , 00000077302YC8 (RU1107CT03); 6375965-203, 00000314588HM5 (RU8157CT01); 6375965-203, 00000342947HM9 (RU8541CT01); 6375965-203, 00000345240HM6 (RU7230CT01); 6375965-203, 00000345242HM2 (RU1192CT01); 6375965-203, 00000381716HM0 (RU8902CT01); 6375965-203, 00000383117HM9 (RU1646CT02); 6375965-203, 00000386889HM0 (RU6973CT01); 6375965-203, 00000388481HM4 (RU9000CT01); 6375965-203, 00000388482HM2 (RU1087CT06); 6375965-203, 00000400972HM6 (RU9564CT01); 6375965-203, 00000405584HM4 (RU1089CT08); 6375965-203, 00000415736HM8 (RU0100CT01); 6375965-203, 00000381717HM8 (RU4574CT01); 6375965-6 , RSGNA1700005YC (RU3138CT03); 6375965-203, 00000345239HM8 (RU3828CT02); 6375965-203, 00000345241HM4 (RU1669CT01); 6375965-203, 00000366680HM7 (RU3720CT01); 6375965-203, 00000385667HM1 (RU1663CT01); 6375965-203, 00000314587HM7 (RU5618CT01); 6375965-6 , RSGNA1700001YC (RU2221CT01); 6375965-6 , RSGNA1800002YC (RU2641CT02); 6375965-203, 00000381714HM5 (RU4573CT01); 6375965-203, 00000386888HM2 (RU1012CT02); 6375965-6 , RSGNA1800001YC (RU1332CT02); 6375965-203, 00000381715HM2 (RU8917CT01); 6375965-6 , RSGNA1800003YC (RU9595CT01); 6375965-6 , RSGNA1700011YC (RU3067CT02); 6375965-203, 00000386523HM5 (RU1786CT03); Model 6375965-123
- 510(k) numbers
- ["K082816","K110227","K131576"]
- PMA numbers
- not on file
- Product code
- JAK
- Firm FEI
- 3002808529
- Firm city
- Hino
- Firm state
- Asahigaoka
- Firm country
- Japan
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28