FDA Medical Device Recalls (openFDA)
In2Bones, SAS: Non-Locking screws in the following lengths: (a) NEO(R) 0.27 - 10 mm, REF W27 ST010 (b) NEO(R) 0.27 - 12 mm, REF W27 ST012 (c) DRP 0.27 - 12 mm, REF W27 ST012 (d) NEO(R) 0.27 - 14 mm, REF W27 ST014 (e) NEO(R) 0.27 - 16 mm, REF W27 ST016 (f) NEO(R) 0.27 - 18 mm, REF W27 ST018 (f) DRP 0.27 - 18 mm, REF W27 ST018 The NEOVIEW(R) Plating System is intended for fixation of intra-articular and extra-articular fractures of the distal radius and reconstruction of the distal radius. It is composed of a NEOVIEW" plate in PEEK and NE0411 screws in Ti6Al4V that are available in Locking or Non-Locking versions. The NEOVIEW411 Screw Set consists of a standard set of NE0411 screws with different lengths suitable for the implantation of a NEOVIEW9 plate for an average anatomy
- Recall number
- Z-2869-2017
- Recalling firm
- In2Bones, SAS
- Product
- Non-Locking screws in the following lengths: (a) NEO(R) 0.27 - 10 mm, REF W27 ST010 (b) NEO(R) 0.27 - 12 mm, REF W27 ST012 (c) DRP 0.27 - 12 mm, REF W27 ST012 (d) NEO(R) 0.27 - 14 mm, REF W27 ST014 (e) NEO(R) 0.27 - 16 mm, REF W27 ST016 (f) NEO(R) 0.27 - 18 mm, REF W27 ST018 (f) DRP 0.27 - 18 mm, REF W27 ST018 The NEOVIEW(R) Plating System is intended for fixation of intra-articular and extra-articular fractures of the distal radius and reconstruction of the distal radius. It is composed of a NEOVIEW" plate in PEEK and NE0411 screws in Ti6Al4V that are available in Locking or Non-Locking versions. The NEOVIEW411 Screw Set consists of a standard set of NE0411 screws with different lengths suitable for the implantation of a NEOVIEW9 plate for an average anatomy
- Status
- Terminated
- Initiated
- 2017-07-13
- Posted
- not on file
- Terminated
- 2018-08-27
- Reason
- The firm received several customer complaints reporting some perioperative breakages of NE0411 Non-Locking screws when inserting the screw into the radius diaphysis
- Root cause
- Device Design
- Action
- In2Bones sent a Field Safety Corrective Action letter dated July 13, 2017, to all affected customers. The consignee was directed to remove the products from distribution and return it to the recalling firm. Customers with questions were instructed to contact the Quality and Regulatory Affairs team at: +33 4 72 29 26 26/ +337 61 88 44 30 or by email: qualite@in2bones.com.
- Quantity
- 287 screws
- Distribution
- Worldwide Diribution - US including Tennessee; Foreign - Italy, Portugal, Spain, Switzerland, UK, and France
- Lot, serial or model codes
- All batch numbers
- 510(k) numbers
- ["K160995"]
- PMA numbers
- not on file
- Product code
- HRS
- Firm FEI
- 3010470577
- Firm city
- Ecully
- Firm state
- not on file
- Firm country
- France
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28