In2Bones, SAS: Non-Locking screws in the following lengths: (a) NEO(R) 0.27 - 10 mm, REF W27 ST010 (b) NEO(R) 0.27 - 12 mm, REF W27 ST012 (c) DRP 0.27 - 12 mm, REF W27 ST012 (d) NEO(R) 0.27 - 14 mm, REF W27 ST014 (e) NEO(R) 0.27 - 16 mm, REF W27 ST016 (f) NEO(R) 0.27 - 18 mm, REF W27 ST018 (f) DRP 0.27 - 18 mm, REF W27 ST018 The NEOVIEW(R) Plating System is intended for fixation of intra-articular and extra-articular fractures of the distal radius and reconstruction of the distal radius. It is composed of a NEOVIEW" plate in PEEK and NE0411 screws in Ti6Al4V that are available in Locking or Non-Locking versions. The NEOVIEW411 Screw Set consists of a standard set of NE0411 screws with different lengths suitable for the implantation of a NEOVIEW9 plate for an average anatomy

Recall number
Z-2869-2017
Recalling firm
In2Bones, SAS
Product
Non-Locking screws in the following lengths: (a) NEO(R) 0.27 - 10 mm, REF W27 ST010 (b) NEO(R) 0.27 - 12 mm, REF W27 ST012 (c) DRP 0.27 - 12 mm, REF W27 ST012 (d) NEO(R) 0.27 - 14 mm, REF W27 ST014 (e) NEO(R) 0.27 - 16 mm, REF W27 ST016 (f) NEO(R) 0.27 - 18 mm, REF W27 ST018 (f) DRP 0.27 - 18 mm, REF W27 ST018 The NEOVIEW(R) Plating System is intended for fixation of intra-articular and extra-articular fractures of the distal radius and reconstruction of the distal radius. It is composed of a NEOVIEW" plate in PEEK and NE0411 screws in Ti6Al4V that are available in Locking or Non-Locking versions. The NEOVIEW411 Screw Set consists of a standard set of NE0411 screws with different lengths suitable for the implantation of a NEOVIEW9 plate for an average anatomy
Status
Terminated
Initiated
2017-07-13
Posted
not on file
Terminated
2018-08-27
Reason
The firm received several customer complaints reporting some perioperative breakages of NE0411 Non-Locking screws when inserting the screw into the radius diaphysis
Root cause
Device Design
Action
In2Bones sent a Field Safety Corrective Action letter dated July 13, 2017, to all affected customers. The consignee was directed to remove the products from distribution and return it to the recalling firm. Customers with questions were instructed to contact the Quality and Regulatory Affairs team at: +33 4 72 29 26 26/ +337 61 88 44 30 or by email: qualite@in2bones.com.
Quantity
287 screws
Distribution
Worldwide Diribution - US including Tennessee; Foreign - Italy, Portugal, Spain, Switzerland, UK, and France
Lot, serial or model codes
All batch numbers
510(k) numbers
["K160995"]
PMA numbers
not on file
Product code
HRS
Firm FEI
3010470577
Firm city
Ecully
Firm state
not on file
Firm country
France

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register