Zimmer Biomet, Inc.: ROSA Brain, 3.0.0 The device is intended for the spatial positioning and orientation of instrument holders or tool guides to be used by neurosurgeons to guide standard neurosurgical instruments (biopsy needle, stimulation or recording electrode, endoscope). The device is indicated for any neurosurgical procedure in which the use of stereotactic surgery may be appropriate.

Recall number
Z-2868-2017
Recalling firm
Zimmer Biomet, Inc.
Product
ROSA Brain, 3.0.0 The device is intended for the spatial positioning and orientation of instrument holders or tool guides to be used by neurosurgeons to guide standard neurosurgical instruments (biopsy needle, stimulation or recording electrode, endoscope). The device is indicated for any neurosurgical procedure in which the use of stereotactic surgery may be appropriate.
Status
Terminated
Initiated
2016-10-11
Posted
not on file
Terminated
2020-06-29
Reason
Complaint of the head holder connector locking up mechanically when tightened.
Root cause
Process design
Action
Product was/will be corrected by Field Technician staff in the form of a visit.
Quantity
20 units (16 US, 4 OUS)
Distribution
Worldwide Distribution - US including CA, FL, MA, MN, NC, NY, OH, PA, DC; Foreign - Australia, and France
Lot, serial or model codes
Device Listing No. D264591, Serial No. BR16004, BR16005, BR16006, BR16009, BR16010, BR16011, BR16012, BR16013, BR16014, BR16015, BR16016, BR16017, BR16018, BR16021, BR16023, BR16024.
510(k) numbers
["K151359"]
PMA numbers
not on file
Product code
HAW
Firm FEI
1000220733
Firm city
Warsaw
Firm state
IN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register