Sentinel CH SPA: MULTIGENT Creatinine (Enzymatic), Catalog Numbers 8L24-31 and 8L24-41; IVD. Intended for the quantitative determination of creatinine in human serum, plasma, and urine using the ARCHITECT c Systems The product is a medical device (chemistry reagent) and is intended for the quantitative determination of creatinine in human serum, plasma, and urine using the ARCHITECT c Systems. The kit is provided as a ready to use liquid material consisting in two components: Reagent 1 and Reagent 2. According to the catalogue number, the kit is composed of three vials of 48 ml for Reagent 1 and three vials of 18 ml for Reagent 2 in the configuration 8L24-31. The configuration 8L24-41 consists in ten vials of 90 ml for Reagent 1 and ten vials of 34 ml for Reagent 2

Recall number
Z-2867-2017
Recalling firm
Sentinel CH SPA
Product
MULTIGENT Creatinine (Enzymatic), Catalog Numbers 8L24-31 and 8L24-41; IVD. Intended for the quantitative determination of creatinine in human serum, plasma, and urine using the ARCHITECT c Systems The product is a medical device (chemistry reagent) and is intended for the quantitative determination of creatinine in human serum, plasma, and urine using the ARCHITECT c Systems. The kit is provided as a ready to use liquid material consisting in two components: Reagent 1 and Reagent 2. According to the catalogue number, the kit is composed of three vials of 48 ml for Reagent 1 and three vials of 18 ml for Reagent 2 in the configuration 8L24-31. The configuration 8L24-41 consists in ten vials of 90 ml for Reagent 1 and ten vials of 34 ml for Reagent 2
Status
Terminated
Initiated
2017-05-22
Posted
2017-06-27
Terminated
2018-09-04
Reason
Reagent 1 component may result in Quality Control values shifting out of acceptable range, which is not resolved by recalibration.
Root cause
Nonconforming Material/Component
Action
Sentinel Diagnostics notified the US Distributor on 5/22/17 to notify US customers by letter. Actions to be Taken 1. Discontinue the use of the impacted lots immediately and destroy any remaining inventory of impacted lots according to your laboratory procedures. 2. If you have an alternate reagent lot available in inventory, switch to this alternate lot, or use a new lot provided. 3. Please review the content of this communication with your Medical Director and retain this letter for any future reference. Questions regarding this information, U.S. Customers should contact Customer Service at 1-877-4 ABBOTT. Customers to complete Field Safety Notice Receipt.
Quantity
2356 kits Worldwide; US 206 kits
Distribution
Nationwide Distribution
Lot, serial or model codes
Catalog Number 8L24-31: Lot Numbers: 60335Y600, 60382Y600 (Exp. Date 2018-02-28); Catalog Number 8L24-41: Lot Number: 60334Y600, (Exp. Date 2018-02-28)
510(k) numbers
["K073634"]
PMA numbers
not on file
Product code
JFY
Firm FEI
3000303338
Firm city
Milan
Firm state
not on file
Firm country
Italy

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register