Philips Medical Systems Gmbh, DMC: DigitalDiagnost, Single-Detect (Stitching Patient Support) 712062

Recall number
Z-2866-2018
Recalling firm
Philips Medical Systems Gmbh, DMC
Product
DigitalDiagnost, Single-Detect (Stitching Patient Support) 712062
Status
Terminated
Initiated
2018-06-27
Posted
2018-08-17
Terminated
2024-04-30
Reason
The hook does not securely hold the footplate in the vertical position. Therefore, the footplate can fall or swing down and injury operator or patient.
Root cause
Device Design
Action
On June 27, 2018, Philips Medical Systems issued URGENT FIELD SAFETY NOTICES, Medical Device Recall letters to customers. Customers were encouraged to always check the right position of the hook and ensure that the hook is locked according tot he IFU. Press hook down until it is locked. Make sure that the hook is in position. A Philips Service Engineer will contact impacted customers as soon as the Field Action Kit is ready to be implemented. Should you need to communicate with Philips in regards to this program, please reference FCO-71200185. If you need further information or support concerning this issue, please contact your local Philips representative.
Quantity
not on file
Distribution
US Nationwide; International to 66 countries
Lot, serial or model codes
Serial Numbers:  386786/SN0702512 375366 416255/SN0902004 358322 388667/SN0802105 375367 397229/SN0802250 396838/SN0802365 386108/SN0802044 405687/SN0802469 354125 362438 406156/SN0802537 413697/SN0802634 402853/SN0902012 405889/SN0802629 383385/SN0802011 378827/SN0802021 384013/SN0702495 358017 358496 377925/SN0802060 356953 355621 357888 356956 368431 356963 406016/SN0802476 346522 375370 70700 351321 392504/SN0802182
510(k) numbers
["K090625"]
PMA numbers
not on file
Product code
MQB
Firm FEI
3003768251
Firm city
Hamburg
Firm state
not on file
Firm country
Germany

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register