Skip to the content

FDA Medical Device Recalls (openFDA)

Sorin Group USA, Inc.: Stckert S5 System, S5 Heart-lung machine, Cardiopulmonary bypass heart-lung machine console Product Usage: The Stckert S5 System is intended to be used during cardiopulmonary bypass for procedures lasting six (6) hours or less. These devices come in direct contact with the central circulatory system but they are not intended to control, diagnose, monitor or correct a defect.

Recall number
Z-2866-2016
Recalling firm
Sorin Group USA, Inc.
Product
Stckert S5 System, S5 Heart-lung machine, Cardiopulmonary bypass heart-lung machine console Product Usage: The Stckert S5 System is intended to be used during cardiopulmonary bypass for procedures lasting six (6) hours or less. These devices come in direct contact with the central circulatory system but they are not intended to control, diagnose, monitor or correct a defect.
Status
Terminated
Initiated
2016-08-25
Posted
2016-09-21
Terminated
2017-05-26
Reason
Sorin/LivaNova is initiating a field correction on the S5 Heart-lung machine because of the potential for failure of the Uninterruptible Power Supply.
Root cause
Employee error
Action
Sorin Group Deutschland GmbH created Field Correction Order (FCO) 2016-003 on August 11, 2016 to replace the affected power supply with a new unit. The customers will be contacted by the Field Service Representatives to schedule the replacement at the customer facilities. No direct action is required by the customer.
Quantity
5264 Units Worldwide
Distribution
Worldwide Distribution - US Nationwide in the states of AZ, CA, CO, FL, HI, KS, MA, NC, TX, UT and the countries of: Great Brittan, Peru, Russia, India, Hungary, Thailand, Germany, Azerbaijan, Austria, Italy, Slovakia, Bahrain, and Malaysia.
Lot, serial or model codes
Item No.: 48-30-00 Serial No: 28E24640, 28E24651, 28E24642, 28E24674, 28E24657, 28E24652, 28E24671, 28E24675, 28E24676, 28E24678, 28E24670  Item No 48-40-00 Serial No: 28E24644, 28E24641, 28E24648, 28E24661, 28E24643, 28E24650, 28E24677, 28E24659, 28E24658, 28E24672, 28E24666, 28E24660, 28E24665, 28E24645, 28E24646, 28E24667, 28E24668, 28E24669, 28E24679  Item No 48-50-00 Serial No: 28E24647, 28E24663, 28E24673, 28E24664  Item No 58-00-00 Serial No.: 28E24653, 28E24654, 28E24655, 28E24656
510(k) numbers
["K071318"]
PMA numbers
not on file
Product code
DTQ
Firm FEI
1718850
Firm city
Arvada
Firm state
CO
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register