Sorin Group USA, Inc.: Stckert S5 System, S5 Heart-lung machine, Cardiopulmonary bypass heart-lung machine console Product Usage: The Stckert S5 System is intended to be used during cardiopulmonary bypass for procedures lasting six (6) hours or less. These devices come in direct contact with the central circulatory system but they are not intended to control, diagnose, monitor or correct a defect.
- Recall number
- Z-2866-2016
- Recalling firm
- Sorin Group USA, Inc.
- Product
- Stckert S5 System, S5 Heart-lung machine, Cardiopulmonary bypass heart-lung machine console Product Usage: The Stckert S5 System is intended to be used during cardiopulmonary bypass for procedures lasting six (6) hours or less. These devices come in direct contact with the central circulatory system but they are not intended to control, diagnose, monitor or correct a defect.
- Status
- Terminated
- Initiated
- 2016-08-25
- Posted
- 2016-09-21
- Terminated
- 2017-05-26
- Reason
- Sorin/LivaNova is initiating a field correction on the S5 Heart-lung machine because of the potential for failure of the Uninterruptible Power Supply.
- Root cause
- Employee error
- Action
- Sorin Group Deutschland GmbH created Field Correction Order (FCO) 2016-003 on August 11, 2016 to replace the affected power supply with a new unit. The customers will be contacted by the Field Service Representatives to schedule the replacement at the customer facilities. No direct action is required by the customer.
- Quantity
- 5264 Units Worldwide
- Distribution
- Worldwide Distribution - US Nationwide in the states of AZ, CA, CO, FL, HI, KS, MA, NC, TX, UT and the countries of: Great Brittan, Peru, Russia, India, Hungary, Thailand, Germany, Azerbaijan, Austria, Italy, Slovakia, Bahrain, and Malaysia.
- Lot, serial or model codes
- Item No.: 48-30-00 Serial No: 28E24640, 28E24651, 28E24642, 28E24674, 28E24657, 28E24652, 28E24671, 28E24675, 28E24676, 28E24678, 28E24670 Item No 48-40-00 Serial No: 28E24644, 28E24641, 28E24648, 28E24661, 28E24643, 28E24650, 28E24677, 28E24659, 28E24658, 28E24672, 28E24666, 28E24660, 28E24665, 28E24645, 28E24646, 28E24667, 28E24668, 28E24669, 28E24679 Item No 48-50-00 Serial No: 28E24647, 28E24663, 28E24673, 28E24664 Item No 58-00-00 Serial No.: 28E24653, 28E24654, 28E24655, 28E24656
- 510(k) numbers
- ["K071318"]
- PMA numbers
- not on file
- Product code
- DTQ
- Firm FEI
- 1718850
- Firm city
- Arvada
- Firm state
- CO
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28