FDA Medical Device Recalls (openFDA)

Philips Medical Systems Gmbh, DMC: DigitalDiagnost, Dual-Detector (Stitching Patient Support) 712052

Recall number
Z-2865-2018
Recalling firm
Philips Medical Systems Gmbh, DMC
Product
DigitalDiagnost, Dual-Detector (Stitching Patient Support) 712052
Status
Terminated
Initiated
2018-06-27
Posted
2018-08-17
Terminated
2024-04-30
Reason
The hook does not securely hold the footplate in the vertical position. Therefore, the footplate can fall or swing down and injury operator or patient.
Root cause
Device Design
Action
On June 27, 2018, Philips Medical Systems issued URGENT FIELD SAFETY NOTICES, Medical Device Recall letters to customers. Customers were encouraged to always check the right position of the hook and ensure that the hook is locked according tot he IFU. Press hook down until it is locked. Make sure that the hook is in position. A Philips Service Engineer will contact impacted customers as soon as the Field Action Kit is ready to be implemented. Should you need to communicate with Philips in regards to this program, please reference FCO-71200185. If you need further information or support concerning this issue, please contact your local Philips representative.
Quantity
not on file
Distribution
US Nationwide; International to 66 countries
Lot, serial or model codes
Serial Numbers:  363679 404164/SN0802462 378830/SN0802004 341693 358494 359442 396571/SN0802311 359449 346075 382021/SN0802313 374374 404298/SN0802474 402570/SN0802420 373501 373500 382272/SN0802113 368428 382012/SN0702520 376667 356957 392502/SN0802179 407423/SN0802638 359767 368619 358641 357697 381998/SN0702488 413829/SN0802631 358320 400364/SN0802522 404166/SN0802441 356981 390272/SN0802364 390273/SN0802366 412482/SN0802683 348997 396576/SN0802233 372848 373508 383084/SN0802127
510(k) numbers
["K090625"]
PMA numbers
not on file
Product code
MQB
Firm FEI
3003768251
Firm city
Hamburg
Firm state
not on file
Firm country
Germany

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register