Philips Medical Systems Gmbh, DMC: DigitalDiagnost Upgrades (Stitching Patient Support) 712083
- Recall number
- Z-2864-2018
- Recalling firm
- Philips Medical Systems Gmbh, DMC
- Product
- DigitalDiagnost Upgrades (Stitching Patient Support) 712083
- Status
- Terminated
- Initiated
- 2018-06-27
- Posted
- 2018-08-17
- Terminated
- 2024-04-30
- Reason
- The hook does not securely hold the footplate in the vertical position. Therefore, the footplate can fall or swing down and injury operator or patient.
- Root cause
- Device Design
- Action
- On June 27, 2018, Philips Medical Systems issued URGENT FIELD SAFETY NOTICES, Medical Device Recall letters to customers. Customers were encouraged to always check the right position of the hook and ensure that the hook is locked according tot he IFU. Press hook down until it is locked. Make sure that the hook is in position. A Philips Service Engineer will contact impacted customers as soon as the Field Action Kit is ready to be implemented. Should you need to communicate with Philips in regards to this program, please reference FCO-71200185. If you need further information or support concerning this issue, please contact your local Philips representative.
- Quantity
- not on file
- Distribution
- US Nationwide; International to 66 countries
- Lot, serial or model codes
- Serial Numbers: 437736/SN09000282 491574/SN12000099 488712 415314/SN0902058 445533/SN10000006 414930/SN08000055 SN13000238 SN13000214 504759/SN12000356 487719/SN11000538 504766/SN12000207 SN14000264 SN14000264 469223/SN10000643 493888/SN12000004 316236 306773 306776 334582 325938 SN08000301 396570/SN0802320 SN08000216
- 510(k) numbers
- ["K090625"]
- PMA numbers
- not on file
- Product code
- MQB
- Firm FEI
- 3003768251
- Firm city
- Hamburg
- Firm state
- not on file
- Firm country
- Germany
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28