Skip to the content

FDA Medical Device Recalls (openFDA)

Philips Medical Systems Gmbh, DMC: DigitalDiagnost Release 3.1 (Stitching Patient Support) 712026

Recall number
Z-2862-2018
Recalling firm
Philips Medical Systems Gmbh, DMC
Product
DigitalDiagnost Release 3.1 (Stitching Patient Support) 712026
Status
Terminated
Initiated
2018-06-27
Posted
2018-08-17
Terminated
2024-04-30
Reason
The hook does not securely hold the footplate in the vertical position. Therefore, the footplate can fall or swing down and injury operator or patient.
Root cause
Device Design
Action
On June 27, 2018, Philips Medical Systems issued URGENT FIELD SAFETY NOTICES, Medical Device Recall letters to customers. Customers were encouraged to always check the right position of the hook and ensure that the hook is locked according tot he IFU. Press hook down until it is locked. Make sure that the hook is in position. A Philips Service Engineer will contact impacted customers as soon as the Field Action Kit is ready to be implemented. Should you need to communicate with Philips in regards to this program, please reference FCO-71200185. If you need further information or support concerning this issue, please contact your local Philips representative.
Quantity
not on file
Distribution
US Nationwide; International to 66 countries
Lot, serial or model codes
Serial Numbers:  SN14000256 SN14000250 SN15000024 SN14000170 SN14000052 SN14000224 SN15000039 SN14000144 SN14000253 SN14000239 SN14000186 SN14000075 SN14000193 SN14000254 SN15000001 SN14000089 SN14000223 SN14000267 SN14000238 SN14000164 SN14000165 SN14000200 SN14000005 SN14000040 SN15000015 SN15000029 SN14000120
510(k) numbers
["K090625"]
PMA numbers
not on file
Product code
MQB
Firm FEI
3003768251
Firm city
Hamburg
Firm state
not on file
Firm country
Germany

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register