Skip to the content

FDA Medical Device Recalls (openFDA)

Globus Medical, Inc.: ALTERA Spacer ALTERA Spacer, 10x31, 12-16mm, 8

Recall number
Z-2861-2020
Recalling firm
Globus Medical, Inc.
Product
ALTERA Spacer ALTERA Spacer, 10x31, 12-16mm, 8
Status
Terminated
Initiated
2020-07-22
Posted
2020-08-22
Terminated
2021-08-19
Reason
Specific lots of ALTERA implants have internal components that may have been manufactured using stainless steel rather than the specified cobalt chromium molybdenum alloy.
Root cause
Nonconforming Material/Component
Action
A customer letter dated July 22, 2020 was mailed to customers. In this letter customers are instructed to immediately examine their inventory and determine if they have any of the affected implants. Alternatively, they may return all ALTERA inventory and Globus will sort out the affected implants and replace inventory with conforming implants. Report any adverse effects believed to be associated with the product to Globus Medical immediately by filling out the attached Response Card.
Quantity
not on file
Distribution
Worldwide Distribution
Lot, serial or model codes
Part #: 1124.1115; Lot # / UDI #: BAX159RE / (01)00889095069716(10)BAX159RE; BAX178JE / (01)00889095069716(10)BAX178JE; BAX208RE / (01)00889095069716(10)BAX208RE; BAX229UE / (01)00889095069716(10)BAX229UE; BAX250BE / (01)00889095069716(10)BAX250BE; BAX308BE / (01)00889095069716(10)BAX308BE; BAX347YE / (01)00889095069716(10)BAX347YE; BAX351KE / (01)00889095069716(10)BAX351KE
510(k) numbers
["K140411"]
PMA numbers
not on file
Product code
MAX
Firm FEI
3004142400
Firm city
Audubon
Firm state
PA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register