Skip to the content

FDA Medical Device Recalls (openFDA)

Philips Medical Systems Gmbh, DMC: DigitalDiagnost Classic, Dual-Detector (Stitching Patient Support) 712052

Recall number
Z-2857-2018
Recalling firm
Philips Medical Systems Gmbh, DMC
Product
DigitalDiagnost Classic, Dual-Detector (Stitching Patient Support) 712052
Status
Terminated
Initiated
2018-06-27
Posted
2018-08-17
Terminated
2024-04-30
Reason
The hook does not securely hold the footplate in the vertical position. Therefore, the footplate can fall or swing down and injury operator or patient.
Root cause
Device Design
Action
On June 27, 2018, Philips Medical Systems issued URGENT FIELD SAFETY NOTICES, Medical Device Recall letters to customers. Customers were encouraged to always check the right position of the hook and ensure that the hook is locked according tot he IFU. Press hook down until it is locked. Make sure that the hook is in position. A Philips Service Engineer will contact impacted customers as soon as the Field Action Kit is ready to be implemented. Should you need to communicate with Philips in regards to this program, please reference FCO-71200185. If you need further information or support concerning this issue, please contact your local Philips representative.
Quantity
not on file
Distribution
US Nationwide; International to 66 countries
Lot, serial or model codes
Serial Numbers:  422102/SN0902020 435210/SN0902204 460471/SN1002019 433651/SN0902114 427610/SN0902108 470160/SN1102003 447013/SN1002008 445778/SN1002006 420208/SN0902019
510(k) numbers
["K090625"]
PMA numbers
not on file
Product code
MQB
Firm FEI
3003768251
Firm city
Hamburg
Firm state
not on file
Firm country
Germany

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register