FDA Medical Device Recalls (openFDA)
EKOS Corporation: EkoSonic Endovascular Devices EKOS Intelligent Drug Delivery Catheter
- Recall number
- Z-2854-2016
- Recalling firm
- EKOS Corporation
- Product
- EkoSonic Endovascular Devices EKOS Intelligent Drug Delivery Catheter
- Status
- Terminated
- Initiated
- 2016-08-09
- Posted
- not on file
- Terminated
- 2016-10-05
- Reason
- The recalling firm identified ten devices which were distributed and could potentially be non-compliant. These devices potentially underwent a routine manufacturing test with non-compliant parameters, which may result in a small blister in the epoxy over transducers in the treatment zone. It is not known at the time the recall was initiated whether the non-compliant test parameters, or presence of the blisters, adversely affect operation of the device.
- Root cause
- Storage
- Action
- All affected consignees were contacted by e-mail on August 9, 2016, and asked to sequester implicated devices.
- Quantity
- 10 units
- Distribution
- Three international consignees - Germany, New Zealand and Taiwan
- Lot, serial or model codes
- Table 1: Device Identification Information Catalog Number; Lot Number; Serial Number; Expiration Date 500-5518; 160711017; 160711017-005; 06-30-2019 500-5518; 160711017; 160711017-006; 06-30-2019. 500-5518; 160711017; 160711017-003; 06-30-2019. 500-5518; 160711017; 160711017-004; 06-30-2019. 500-5512; 160711042; 160711042-010; 06-30-2019. 500-5518; 160711017; 160711017-001; 06-30-2019. 500-5518; 160711017; 160711017-002; 06-30-2019. 500-5518; 160711017; 160711017-007; 06-30-2019. 500-5518; 160711017; 160711017-008; 06-30-2019. 500-5518; 160711017; 160711017-009; 06-30-2019. 3001627457 08/09/2016-002R CONFIDENTIAL 3 ¿ Name, address, and type of responsible firm on label; number and description of private labels. Manufacturer: EKOS Corporation 11911 North Creek Parkway South Bothell, WA 98011 ¿ Collect a complete copy of all labeling (include product inserts, information sheets.) Applicable labeling is included in Attachment 1. B. CODE: ALL lot and/or serial numbers, batch numbers, catalog numbers, product numbers, packer or manufacturer numbers, etc., exactly as they appear on the product, container, or labeling. There are 10 total devices in distribution. Identification information for the implicated devices is found in Table 1: Table 1: Device Identification Information Catalog Number Lot Number Serial Number Expiration Date 500-5518 160711017 160711017-005 06-30-2019 500-5518 160711017 160711017-006 06-30-2019 500-5518 160711017 160711017-003 06-30-2019 500-5518 160711017 160711017-004 06-30-2019 500-5512 160711042 160711042-010 06-30-2019 500-5518 160711017 160711017-001 06-30-2019 500-5518 160711017 160711017-002 06-30-2019 500-5518 160711017 160711017-007 06-30-2019 500-5518 160711017 160711017-008 06-30-2019 500-5518 160711017 160711017-009 06-30-2019 ¿ Give a breakdown of the code Catalog Number consists of 7 digits (XXX-YYYY), where; ¿ XXX represents a product family, ¿ YYYY represents a specific configuration, such as working length/treatment zone length.
- 510(k) numbers
- ["K081467","K111705","K140151"]
- PMA numbers
- not on file
- Product code
- KRA
- Firm FEI
- 3001627457
- Firm city
- Bothell
- Firm state
- WA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28