Skip to the content

FDA Medical Device Recalls (openFDA)

INTEGRATED MEDCRAFT LLC: FORWARD Oral Airway Kit (OPA Sizes 5, 6.5, 7, 8, 9, 10, 12). REF Number: 1011-40002. Containing the affected One-piece Guedel airway components: 1. Size 00, ISO 5.0, blue. Product Code: 1100050. 2. Size 000, ISO 3.5, light green. Product Code: 1000035.

Recall number
Z-2836-2026
Recalling firm
INTEGRATED MEDCRAFT LLC
Product
FORWARD Oral Airway Kit (OPA Sizes 5, 6.5, 7, 8, 9, 10, 12). REF Number: 1011-40002. Containing the affected One-piece Guedel airway components: 1. Size 00, ISO 5.0, blue. Product Code: 1100050. 2. Size 000, ISO 3.5, light green. Product Code: 1000035.
Status
Open, Classified
Initiated
2026-07-07
Posted
2026-07-31
Terminated
not on file
Reason
Devices could potentially contain an occluded or partially occluded Guedel Airway tip.
Root cause
Component design/selection
Action
Firm initially notified customers on June 23, 2026 via email and June 24,2026 via postal mail. Updated letters were sent on July 7, 2026 via email. Customers were instructed to immediately discontinue use of affected product and inspect inventory for affected product. Customers should return affected product to Integrated MedCraft. If the customer has further distributed affected product, they should notify customers who received the product.
Quantity
509 units
Distribution
US Nationwide distribution.
Lot, serial or model codes
FORWARD Oral Airway Kit (OPA Sizes 5, 6.5, 7, 8, 9, 10, 12). REF Number: 1011-40002. Lot Numbers: IMC-00000060, IMC-00000066, IMC-00000078, IMC-00000086, IMC-00000092, IMC-00000096, IMC-00000096, IMC-00000097, IMC-00000103, IMC-00000110, IMC-00000125, IMC-00000151, IMC-00000200, IMC-00000201, IMC-00000204, IMC-00000205. Containing the affected One-piece Guedel airway components: 1. Size 00, ISO 5.0, blue. Product Code: 1100050. Lot Numbers: 32121232, 32208193, 32400267, 32403734, 32408819, 32409672, 32416545, 32427064, 32605471, 32606405, 32606850, 32607154. 2. Size 000, ISO 3.5, light green. Product Code: 1000035. Lot Numbers: 32190425, 32213076, 32221283, 32307261, 32317627, 32321305, 32513841, 32517565
510(k) numbers
not on file
PMA numbers
not on file
Product code
CAE
Firm FEI
3030078769
Firm city
East Hampton
Firm state
CT
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register