FDA Medical Device Recalls (openFDA)

LIEBEL-FLARSHEIM COMPANY LLC: Liebel-Flarsheim Hydra Vision Urology X-Ray System (DR) Liebel-Flarsheim urology systems facilitate radiologic and/or fluoroscopic procedures requiring a beam of diagnostic quality radiation, primarily for urological applications such as functional x-ray diagnostics, endourology and minimal invasive urology/ surgery.

Recall number
Z-2833-2018
Recalling firm
LIEBEL-FLARSHEIM COMPANY LLC
Product
Liebel-Flarsheim Hydra Vision Urology X-Ray System (DR) Liebel-Flarsheim urology systems facilitate radiologic and/or fluoroscopic procedures requiring a beam of diagnostic quality radiation, primarily for urological applications such as functional x-ray diagnostics, endourology and minimal invasive urology/ surgery.
Status
Terminated
Initiated
2017-02-27
Posted
not on file
Terminated
2023-08-16
Reason
Software issue. The difference between the display and dosimeter readings In the Child/Pediatric automatic exposure mode is in the range of 52-65%. The display in all other modes deviated by less than 35% from the dosimeter readings.
Root cause
Radiation Control for Health and Safety Act
Action
The firm, Liebel-Flarsheim, notified its customers by phone on or about 2/27/2017 to inform them of the product, problem and actions to be taken. The firm plans to have these Sedecal software updates implemented over the course of the next 12 months. For additional information, please contact me at 314-376-4788.
Quantity
313
Distribution
US Distribution to: AL, AZ, CA, CT, CO, FL GA, HI, IA, IL, ID, KS, KY, MD, MI, Michigan, MN, NC, NH, NV, NY, OK, OH, PA, SC, TN, TX, VA, WA, WI, and WV,
Lot, serial or model codes
Product ID : 750732 SHF-635RF, 750733 SHF-835RF
510(k) numbers
["K943581"]
PMA numbers
not on file
Product code
IXR
Firm FEI
3012465398
Firm city
Saint Louis
Firm state
MO
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register