Skip to the content

FDA Medical Device Recalls (openFDA)

Cytocell Ltd.: Cytocell LPH533-A NUP98 Distal Probe Green - Product Usage:The NUP98 Distal probe, labelled in green, consists of a 151kb probe distal to the NUP98 gene, covering the ZNF195 gene and the D11S4731 marker. Fluorescence in situ hybridisation (FISH) is a technique that allows the visualisation of DNA sequences upon chromosomes.

Recall number
Z-2829-2020
Recalling firm
Cytocell Ltd.
Product
Cytocell LPH533-A NUP98 Distal Probe Green - Product Usage:The NUP98 Distal probe, labelled in green, consists of a 151kb probe distal to the NUP98 gene, covering the ZNF195 gene and the D11S4731 marker. Fluorescence in situ hybridisation (FISH) is a technique that allows the visualisation of DNA sequences upon chromosomes.
Status
Terminated
Initiated
2020-07-17
Posted
not on file
Terminated
2020-09-30
Reason
A low risk of a false positive result being issued with a laboratory developed test (LDT) that utilizes the LPH533-A NUP98 Distal Probe Green.
Root cause
Under Investigation by firm
Action
Cytocell issued Urgent Medical Device Recall Letter dated July 17,200 stating reason for recall, health risk and action to take: Immediately examine your inventory and quarantine all product subject to recall. Cytocell requests that you destroy the remaining inventory. Please complete and return the enclosed response form as soon as possible. We wish to sincerely apologise for any inconvenience caused as a result of this medical device recall. If you have any questions, contact Cytocell on Tel: +44 (0)1223 294048, Fax +44(0)1223 294986 or vigilance@ogt.com.
Quantity
62 units
Distribution
Worldwide distribution - US Nationwide distribution and the country of Canada.
Lot, serial or model codes
Lot numbers: 069915, 070433, 069305, 069468
510(k) numbers
not on file
PMA numbers
not on file
Product code
MVU
Firm FEI
3002646096
Firm city
Cambridge
Firm state
not on file
Firm country
United Kingdom

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register